For research use only. NOT for use in diagnostic procedures.
Figure 1. Structure of 11-nor-9-carboxy-Δ9-THC.
This application note describes the fully automated extraction of carboxy-THC from urine, following base hydrolysis prior to GC/MS analysis. The method was automated using Biotage® ExtraheraTM , configured for use with EVOLUTE EXPRESS ABN cartridges.
The method delivers clean extracts and analyte recovery of 90% with RSD lower than 10% for carboxy-THC and its deuterated internal standard. Using the Biotage® Extrahera, 24 samples were extracted in approximately 38 minutes.
The limit of quantitation of 2 ng/mL is below the SAMHSA/ EWDTS confirmation cut off of 15 ng/mL for workplace testing applications.
The EVOLUTE® EXPRESS range of SPE cartridges and 96-well plates combine powerful EVOLUTE sorbent chemistry with innovative features that provide robust, reliable SPE, giving ultra-pure extracts and high analyte recoveries.
With improved flow-through characteristics and by eliminating the need for conditioning and equilibration, samples can be prepared much faster using the simplified Load-Wash-Elute procedure.
Carboxy-THC and carboxy-THC-D9 as internal standard
EVOLUTE® EXPRESS ABN 60 mg/3 mL tabless SPE cartridges, part number 610-0006-BXG.
Apply 40 μL of a 1 ng/μL aqueous internal standard solution to 2 mL of urine, and allow to equilibrate for 1 hr at room temperature. Add 100 µL sodium hydroxide (10N) to this urine sample and heat at 60 °C for 20 minutes. Allow to cool and add 1 mL glacial acetic acid.
Load the full sample volume onto the column and apply positive pressure to achieve approximately 3 mL/minute. Gravity or very low pressure may be sufficient for this step.
Apply 2 mL of HPLC water and apply pressure to elute.
Apply a further 2 mL of HPLC water and apply pressure to elute.
Dry the column for 10 mins using the Biotage® ExtraheraTM dry plate function, or at 10 psi using other processing systems.
Elute direct into GC vials with hexane/ethyl acetate (1 mL, 50/50, v/v)
Dry the extract in a stream of air or nitrogen using a TurboVap (1.2 bar at 40 °C for 10 mins).
Upon dryness, reconstitute with 20 μL ethyl acetate and 20 μL BSTFA:TMCS 99:1 and vortex for 20 seconds. Place in a heating block set to 70 oC, for 25 minutes. Remove vial from the block and allow to cool.
Agilent 7890A with QuickSwap
Restek Rxi-5ms, 30 m x 0.25 mm ID x 0.25 μm
Helium 1.2 mL/min (constant flow)
280 °C, Splitless, purge flow: 50 mL/min at 1.0 min
2 µL
Methanol and ethyl acetate
Initial temperature 125 °C
Ramp 50 °C/min to 300 °C, hold for 2.5 minutes
Ramp 50 °C/min to 330 °C, hold for 1.4 minutes
Backflush for 1.6 minutes (2 void volumes)
280 °C
Agilent 5975C
230 °C
150 °C
SIM
Table 1. Ions acquired in the Selected Ion Monitoring (SIM) mode.
|
SIM Group |
Analyte |
Target (Quant) Ion |
1st Qual Ion |
2nd Qual Ion |
|
1 |
THC-COOH-D9 |
380 |
479 |
|
|
1 |
THC-COOH |
371 |
488 |
473 |
This optimized EVOLUTE® EXPRESS ABN protocol demonstrated analyte recoveries of 90% and 92% from urine for the carboxy- THC-D9 and carboxy-THC respectively, as shown in Figure 2. RSDs were lower than 10% (n=7).
Figure 3. A calibration curve of carboxy-THC constructed following extractions using EVOLUTE® EXPRESS ABN with the optimal protocol. Analyte concentrations visible here are 2, 5, 10, 20, 50 and 100 ng/ mL showing r2 values of greater than 0.999. The carboxy-d9 deuterated internal standard concentrations is at 20 ng/mL.
Signal to noise measurement shows LOQ to be 2 ng/mL.
|
Part Number |
Description |
Quantity |
|
610-0006-BXG |
EVOLUTE® EXPRESS ABN 60 mg/3 mL cartridge |
50 |
|
414001 |
Biotage® ExtraheraTM |
1 |
|
415041 |
Configuration Kit 24 Positions, Dual Flow |
1 |
|
415000 |
TurboVap® LV Evaporator |
1 |
Literature number: AN893