Biotage Blog

Does dry silica reduce recovery during C18 SPE cleanup?

Written by Symone Stribling | Nov 24, 2026, 8:00:00 AM

New lead compounds are often identified from diverse libraries generated through rapid, small-scale synthesis and purification workflows. However, purification can become a bottleneck before screening, especially for peptides and peptidomimetics that require reversed-phase conditions. Chemists must therefore balance speed, reproducibility, and cost when building compound libraries. To help accelerate peptide and peptoid cleanup, Biotage developed Biotage® PeptiRen-96, a C18-based 96-well SPE plate designed to reduce purification time and solvent consumption by up to 98% compared with traditional approaches.

A previous Biotage blog examined the effect of increasing acetonitrile concentration on purity and recovery during SPE cleanup with Biotage® PeptiRen-96 on the Biotage® Extrahera™ and found that increasing the mid-step MeCN concentration from 24% to 30% improved both recovery and purity. But what happens when that same optimized gradient is applied to dry silica-based C18 media? In purification workflows, it is widely understood that silica-based media should not be allowed to dry during use, as drying can lead to channeling, reduced surface interaction with the sample, and premature elution of the target compound. Using the Biotage® Extrahera™ and Biotage® PeptiRen-96, the effect of dry silica on recovery and purity was evaluated. Peptoid N(FkF) (figure 1) was selected as the test compound due to its aromaticity and relevance to early-stage discovery workflows. The same 30% step gradient from the previous blog was used: conditioning with 100% MeCN containing 0.1% TFA, followed by equilibration with 100% H2O containing 0.1% TFA, sample loading in methanol, wash 1 at 100% H2O containing 0.1% TFA, wash 2 at 5% MeCN containing 0.1% TFA, elution 1 at 30% MeCN containing 0.1% TFA, and elution 2 (flush) at 70% MeCN containing 0.1% TFA.

How the dry-media effect was tested

 

Figure 1: Structure of peptoid N (FkF)

 

Table 1. SPE method used on Biotage® PeptiRen-96.

Step

Gradient B

Condition

100% MeCN, 0.1% TFA

Equilibrate

100% H2O, 0.1% TFA

Load

Sample in MeOH

Wash 1

100% H2O, 0.1% TFA

Wash 2

5% MeCN, 0.1% TFA

Elution 1

30 % MeCN, 0.1% TFA

Elution 2 (flush)

70% MeCN, 0.1% TFA

What happened when the media dried slightly

In the first experiment, one air bubble was intentionally introduced between each method step, including conditioning, equilibration, both washes, and both elutions, to create slight drying of the media. Liquid was expelled from the full media bed until a single air bubble was observed. Under normal operating conditions, the method leaves a small volume of solvent above the PeptiRen resin bed to keep the media fully wetted throughout the workflow. This approach was used to determine whether even minimal drying of the resin bed affects performance. A total of 9.61 mg of N(FkF) was loaded in 375 µL of MeOH, and the wash and elution fractions were collected. The 0%, 5%, 24% or 30%, and 70% MeCN fractions were then analyzed for mass recovery and purity, with the results summarized in table 2.

Table 2. Recovery and purity results for one air bubble.

Step

1 air bubble recovery (mg) and purity (%)

Wash 1

6.9 mg, 72%; no separation observed, 51%

Wash 2

-

Elution 1

0.8 mg, 8%; 97%

Elution 2 (flush)

-

 

Most of the sample eluted during the 100% H2O wash 1 step, suggesting that media dryness reduced retention. Only a small portion of the target compound was recovered in the intended 30% MeCN fraction.

Effect of fully dried silica on peptide purification

To confirm these findings, a second experiment was performed in which the silica media was fully dried between each step after sample loading. Complete drying was achieved by applying five times the standard pressure, allowing the same volume to pass through over more than twice the usual duration. Peptoid N(FkF) was used again, with 10.1 mg dissolved in 375 µL of MeOH, and the elution fractions were collected.

The results showed:

  • Wash 2 at 5% MeCN: 0.4 mg, consisting entirely of impurities

  • Elution 1 at 30% MeCN: 0.9 mg at 84.1% purity

  • Elution 2 at 70% MeCN: 0.5 mg, consisting entirely of impurities

Compared with the first experiment, recovery and purity in the 30% MeCN fraction decreased further, and material was again absent from the expected elution points. The data are summarized in table 3.

Table 3: Recovery and purity results for dry media across wash and elution steps when using Biotage® PeptiRen-96 on the Biotage® Extrahera Classic.

Step

Dry media recovery (%) and purity (%)

Wash 2

-

Elution 1

8%; 84%

Elution 2 (flush)

 -

 

Why keeping the media wet matters

Together, these experiments show what many purification scientists recognize in practice but rarely quantify: dry silica can reduce sample retention and compromise separation performance. In both cases, premature elution occurred during the 0% MeCN wash step, and target-compound recovery decreased despite use of the optimized MeCN gradient. By limiting sample-to-surface interaction, dry media can weaken binding and reduce purification efficiency.

Best practices for maintaining C18 SPE recovery

Allowing C18 SPE wells to dry between steps—whether through air gaps or complete drying—reduced both recovery and purity during crude sample cleanup. Even under optimized elution conditions, dry media negatively affected retention of peptoid N(FkF). These findings reinforce a simple but important best practice: keep the media wet. Maintaining hydration throughout the SPE workflow promotes effective binding and more reproducible purification.

Interested in learning more about solid-phase extraction cleanup strategies for peptides and peptoids?

 