Figure 1. Structure of Butabarbital
This application note describes the extraction of barbiturate compounds from whole blood, prior to GC/MS analysis. This protocol also allows the simultaneous extraction of various other drugs of abuse classes: amphetamines, benzodiazepines, cocaine and opiates.
ISOLUTE® SLE+ cartridges with 1 mL sample capacity are used to extract whole blood samples following a straightforward sample dilution. No protein precipitation or other pre-treatment is required prior to sample loading. The sample preparation procedure delivers clean extracts, good recoveries and RSD values and LLOQs from 20 ng/mL (analyte dependant).
ISOLUTE SLE+ Supported Liquid Extraction plates and cartridges offer an efficient alternative to traditional liquid- liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation, and significantly reduced sample preparation.
Butalbarbital, Butabarbital, Amobarbital, Pentobarbital, Secobarbital, Hexobarbital, Phenobarbital
ISOLUTE® SLE+ 1mL Sample Volume cartridge, part number 820-0140-C
To 1 mL of whole blood, add 10 µL of ISTD (total 100 ng/mL). Allow to equilibrate and add 1 mL of 1% ammonium hydroxide (aq). Vortex.
Load 750 μL of the pre-treated whole blood onto the cartridge and apply a pulse of vacuum or positive pressure (3–5 seconds) to initiate flow. Allow the sample to absorb for 5 minutes
Apply dichloromethane* (DCM, 2.5 mL) and allow to flow under gravity for 5 minutes. Collect in an appropriate glass tube. Apply a second aliquot of DCM* (2.5 mL) and allow to flow under gravity for 5 minutes. Apply vacuum or positive pressure (5-10 seconds) to pull through any remaining extraction solvent into the collection vessel.
* Note that MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, benzodiazepines, opiates). If simultaneous extraction of cocaine and metabolites is required, DCM should be used as extraction solvent.
Evaporate the extract in a stream of air or nitrogen using a TurboVap® LV (40 °C, 20 to 40 L/min). Reconstitute the extracts with ethyl acetate (250 µL) and vortex for 20 seconds before transferring to high recovery GC vials. Evaporate the extract in a stream of air or nitrogen using a SPE Dry (40 °C, 20 to 40 L/min). Reconstitute extracts with ethyl acetate (80 µL) and TMAH (trimethylanilinum hydroxide) (20 µL) and vortex.
Agilent 7890A with QuickSwap
Agilent J&W DB-5, 30 m x 0.25 mm ID x 0.25 μm
Helium 1.2 mL/min (constant flow)
150 °C, Splitless, purge flow: 50 mL/min at 1.0 min
1 µL
Acetone & ethyl acetate
Initial temperature 120 °C, hold for 1 minute, ramp 12 °C/min to 192 °C, ramp 100 °C/min to 330 °C
Backflush for 1.6 minutes (2 void volumes)
280 °C
Agilent 5975C
230 °C
150 °C
SIM
|
SIM Group |
Analyte |
Target (Quant) Ion |
Qual Ion 1 |
Qual Ion 2 |
|---|---|---|---|---|
|
1 |
Butalbital |
196 |
181 |
195 |
|
1 |
Butalbital |
169 |
184 |
211 |
|
2 |
Amobarbital |
169 |
184 |
226 |
|
3 |
Pentobarbital |
169 |
112 |
184 |
|
4 |
Secobarbital |
196 |
181 |
195 |
|
5 |
Hexobarbital |
235 |
79 |
169 |
|
5 |
Phenobarbital |
232 |
117 |
175 |
Blank whole blood was spiked at 100 ng/mL for recovery testing; the typical recovery data is shown in Figure 2. Both MTBE and DCM protocols gave reproducible data with RSD values <10%.
Figure 2. Typical opiate recoveries using MTBE or DCM as extraction solvent.
extraction protocol.
Whole blood was spiked prior to extraction, at concentrations of 10, 20, 50, 75, 100, 200 and 500 ng/mL for each analyte to create calibration curves. Morphine-D3 and 6-MAM-D3 were spiked at 100 ng/mL for each level. The curves are shown in Figure 4.
|
Drug Analyte |
LLOQ (ng/mL) DCM Extraction |
LLOQ (ng/mL) MTBE Extraction |
|
Butalbarbital |
50 |
50 |
|
Butabarbital |
20 |
20 |
|
Amobarbital |
20 |
20 |
|
Pentobarbital |
20 |
50 |
|
Secobarbital |
50 |
50 |
|
Hexobarbital |
20 |
20 |
|
Phenobarbital |
100 |
20 |
ISOLUTE® SLE+ cartridges are underloaded (750 µL sample on a 1 mL capacity cartridge) to avoid breakthrough of whole blood matrix.
MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, benzodiazepines, opiates), but cannot be used for cocaine and metabolites.
|
Part Number |
Description |
Quantity |
|
820-0140-C |
ISOLUTE® SLE+ 1 mL Sample Volume Cartridge* |
30 |
|
820-0140-CG |
ISOLUTE SLE+ 1 mL Sample Volume Cartridge (tabless) |
30 |
|
PPM-48 |
Biotage® PRESSURE+ 48 Positive Pressure Manifold |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry Sample Concentrator System 220/240 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry Sample Concentrator System 100/120 V |
1 |
Literature number: AN853