This application note describes the extraction of a panel of 18 steroid hormones from horse hair using ISOLUTE® SLE+ Supported Liquid Extraction plates prior to LC/MS-MS analysis. The simple sample preparation procedure delivers clean extracts and analyte recoveries greater than 80% with RSDs lower than 5% for most analytes. Linearity of greater than 0.999 is achieved for all analytes in the range 0.5-500 pg/mg of hair.
ISOLUTE® SLE+ Supported Liquid Extraction plates offer an efficient alternative to traditional liquid-liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation, and significantly reduced sample preparation.
21-Deoxycortisol, Cortisone, Estradiol, Aldosterone, 11-Deoxycortisol, Corticosterone, Estrone, Dehydroepiandrosterone (DHEA), 17-OH-Progesterone, Testosterone, Dihydrotestosterone (DHT), Androstenedione, 11-deoxycorticosterone, Progesterone, α-trenbolone, β-trenbolone, α-boldenone, β-boldenone.
Dihydrotestosterone- D3 (DHT-D3) and Aldosterone-D4.
ISOLUTE® SLE+ 400 µL sample capacity plate, (p/n) 820-0400-P01
Weigh 20 mg of hair into 2 mL Biotage® Lysera tubes containing 4 x 2.8 mm stainless steel beads.
Grind the hair sample using Biotage® Lysera (conditions: 5 x 6.45 m/s for 45 sec with a 15 sec dwell).
Add 1 mL of 0.1% (v/v) NH4OH in propan-2-ol (IPA) to each ground hair sample. Spike with 100 µL of a 10 pg/µL methanolic ISTD to give a final concentration of 50 pg/mg of hair.
To ensure complete transfer of the analytes into the extraction solvent, process the sample again using Biotage® Lysera (conditions: 3 x 5.3 m/s for 60 sec with a 20 sec dwell).
Centrifuge tubes for 10 minutes at 13,300 rpm.
Transfer 200 µL of extract and load directly onto the ISOLUTE® SLE+ 400 µL plate.
Load up to 200 µL of horse hair extract into each ISOLUTE® SLE+ well. Using a Biotage® PRESSURE+96 Positive Pressure Manifold, apply a pulse of pressure to load samples onto the sorbent. Wait 5 minutes for the sample to equilibrate on the sorbent.
Apply an aliquot of Dichloromethane (DCM) (500 µL) and allow to flow under gravity for 5 minutes. Apply a further aliquot of DCM (500 µL) and allow to flow under gravity for 5 minutes. For complete removal apply a pulse of positive pressure at 10 psi (10–20 seconds). Collect extracts into a 2 mL, 96-well collection plate.
Evaporate extracts at 40 °C, for 30 mins at a flow rate of 20–40 L/min using the Biotage® SPE Dry 96. Reconstitute extracts in a mix of mobile phase A/mobile phase B (50:50, v/v, 200 µL). Vortex mix. Cover plate with a sealing mat prior to injection.
Shimadzu Nexera X2 UHPLC
ACE C18 (100 mm x 2.1 mm, 1.7 µm) (Hichrom (VWR), UK ) with EXP Guard column holder fitted with a C-18 cartridge (Thames Restek, UK)
A: 0.2 mM Ammonium Fluoride (aq)
B: Methanol
0.4 mL/min
40 °C
5 µL
Table 1. UHPLC gradient.
|
Time (min) |
%A |
%B |
Column Oven |
|
0.1 |
Divert to waste |
||
|
2 |
50 |
50 |
|
|
3 |
To MS |
||
|
5 |
40 |
60 |
|
|
8 |
10 |
90 |
|
|
9 |
5 |
95 |
|
|
9.1 |
5 |
95 |
|
|
9.2 |
50 |
50 |
Shimadzu 8060 Triple Quadrupole MS using ES interface
3 L/min
3 L/min
17 L/min
400 oC
250 oC
400 oC
400 oC
270 kPa
For optimum sensitivity, data was acquired in both positive and negative ion modes, as appropriate, shown in Table 2.
Table 2. MS conditions and retention times for target analytes in positive and negative mode.
This simple sample preparation method delivers clean extracts and analyte recoveries greater than 80% with RSDs lower than 5% for most analytes (see Figure 2), and LLOQs as low as 1 pg/mg for the steroids. Figure 2. below shows recoveries using 400 µL capacity ISOLUTE® SLE+ plates, loading 200 µL of non-aqueous extract directly onto the plate.
Figure 3. demonstrates representative chromatography obtained from horsehair spiked at 50 pg/mg. Satisfactory resolution of the various isobars was obtained using the ACE C18 UHPLC column. In order to achieve low level detection of analytes in positive and negative ion modes a combination of 0.2 mM NH4F (aq) and MeOH was utilized in the mobile phase.
Calibration curve performance was investigated using hair from a female horse spiked between 0.5–500 pg/mg. Good linearity was observed for all analytes typically delivering r2 values greater than 0.999. Table 3. details linearity performance and associated LOQ for each analyte loading 200 µL of horse hair extract. Selected calibration curves loading 200 µL are shown in Figure 4.
Table 3. Analyte calibration curve r2 and LOQ performance.
|
Analytes |
r2 |
LLOQ (pg/mg) |
|
β-trenbolone |
0.9993 |
<5 |
|
α-trenbolone |
0.9997 |
0.5 |
|
β-boldenone |
0.9995 |
<0.5 |
|
Cortisone |
0.9997 |
1 |
|
21-Deoxycortisol |
0.9996 |
<5 |
|
Estradiol |
0.9995 |
<10 |
|
Aldosterone |
0.9995 |
25 |
|
α-boldenone |
0.9993 |
<5 |
|
11-Deoxycortisol |
0.9995 |
0.5 |
|
Corticosterone |
0.9995 |
<1 |
|
Estrone |
0.9994 |
0.5 |
|
11-Deoxycorticosterone |
0.9994 |
5 |
|
DHEA |
0.9991 |
25 |
|
Testosterone |
0.9991 |
1 |
|
DHT |
0.9995 |
<5 |
|
Androstenedione |
0.9994 |
1 |
|
17-OH-Progesterone |
0.9995 |
10 |
|
Progesterone |
0.9991 |
<25* |
*Progesterone LOQ was estimated from endogenous levels of a female horse based on peak height and blank baseline. Endogenous levels were approximately 250 pg/mg.
|
Part Number |
Description |
Quantity |
|
19-060 |
Biotage® Lysera |
1 |
|
19-649 |
2 mL Reinforced Tubes with screw caps (Bulk pack) |
1000 |
|
19-640 |
2.4 mm Metal Beads - 500 grams |
1 |
|
820-0055-B |
ISOLUTE® SLE+ 400 µL Sample Volume Columns |
50 |
|
820-0400-P01 |
ISOLUTE® SLE+ 400 µL Capacity Plate |
1 |
|
PPM-96 |
Biotage® PRESSURE+ 96 Positive Pressure Manifold |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry 96 Sample Evaporator 220/240 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry 96 Sample Evaporator 100/120 V |
1 |
|
121-5203 |
Collection Plate, 2 mL Square |
50 |
|
121-5204 |
Piercable Sealing Mat |
50 |
Literature Number: AN921