This method is applicable to the determination of Oil and Grease as defined by EPA Method 1664 as Hexane Extractable Material (HEM, also referred to as Total Oil and Grease) as well as Silica Gel Treated- Hexane Extractable Material (SGT-HEM, also referred to as Total Petroleum Hydrocarbons or TPH).
Oil and Grease samples contain a range of polar and non-polar compounds. Using ISOLUTE® O&G SPE cartridges, both polar and non-polar compounds are extracted from water samples.
The compounds can be fractionated into the non-polar total petroleum hydrocarbon (TPH or SGT-HEM) fraction and remaining polar fraction using a dual elution solvent strategy. If fractionation is not required, a single elution, eluting total Oil & Grease (or HEM) is used. O&G or TPH levels are determined gravimetrically after gentle evaporation of the elution solvent.
ISOLUTE® O&G solid phase extraction cartridges are available in two sizes, and contain an integral depth filter which prevents blocking by highly particulate laden or contaminated waste- water samples.
ISOLUTE® O & G ISOLUTE® O&G 1 g/6 mL Cartridge (Depth Filter) Part Number 753-0100-CD or ISOLUTE® O & G ISOLUTE® O&G 3 g/70 mL Cartridge (Depth Filter) Part Number 753-0300-FD
Acidify the sample to pH of 1.9 to 2.1 with 6 M HCl to neutralize fatty acids. Add 10 mL of methanol to 1 L of sample. Chill to 4 ˚C for samples with high concentrations of fatty acids.
Rinse the extraction cartridge with 10 mL of methanol at 10 mL/min.
Rinse the extraction cartridge with 10 mL of reagent water, acidified to pH between 1.9 and 2.1 with 6 M HCl, at 10 mL/min.
The sample may be loaded at rates not exceeding 10 mL/min. After the method has been optimized, increased loading rates should be tested. Loading rates between 50 and 100 mL/min are realistic.
Rinse the sample bottle with 20 mL of reagent water acidified to pH ~2 with 6 M HCl. If high concentrations of fatty acids are present, repeat this rinse step. Load onto the extraction cartridge. Add 10 mL acetone to the sample bottle. Shake well, making sure that the sides come into good contact with the acetone. Dilute the acetone with 40 mL of reagent water acidified to pH ~2 with 6 M HCl. Swirl. Load onto the extraction cartridge. Repeat this acetone rinse until the bottle washings appear clear. Dry the extraction cartridge for approximately 30 minutes on a vacuum manifold or with gas (nitrogen or carbon dioxide at 4 L/min). When dry, the cartridge will no longer feel cool to the touch.
|
Part Number |
Description |
Quantity |
|
753-0100-CD |
ISOLUTE® O&G 1 g/6 mL Cartridge (Depth Filter) |
30 |
|
753-0300-FD |
ISOLUTE® O&G 3 g/70 mL Cartridge (Depth Filter) |
16 |