Literature

Fast extraction of basic drugs from urine with no dry down step using EVOLUTE® EXPRESS CX mixed-mode cation exchange SPE

Written by Biotage | Mar 5, 2026 3:40:09 PM

Introduction

The screening of patient urine samples for basic drugs is typically performed following a lengthy sample preparation methodology that requires a time consuming dry down step. A novel elution protocol coupled to the EVOLUTE EXPRESS advanced plate technology affords a fast sample preparation solution that follows a LOAD-WASH-ELUTE-ANALYZE strategy. Basic drugs are fortified into urine and extracted in 3 easy steps using EVOLUTE EXPRESS CX solid phase extraction in a 96-well plate format.


Figure 1. Structure of Diazepam, Fentanyl, and Codeine.

EVOLUTE EXPRESS CX mixed-mode cation exchange polymer-based SPE sorbent extracts a wide range of basic analytes from biological fluids and other aqueous matrices using new plate design technology that reduces processing time through the elimination of the traditional conditioning and equilibration steps. EVOLUTE EXPRESS CX removes matrix components such as proteins, salts, non-ionizable interferences and phospholipids, delivering cleaner extracts with reproducible recoveries for accurate quantitation.

Analytes

Alprazolam, Clonazepam, Diazepam, Flunitrazepam, Oxazepam, Temazepam, Nitrazepam, Normeperdine, Naltrexone, Morphine, Codeine, Oxymorphone, Hydromorphone, Oxycodone, Hydrocodone, 6-Acetyl Codeine, 6-Acetyl Morphine, Fentanyl, Buprenorphine, EDDP

Sample preparation workflow:

Load >> Wash >> Elute >> Analyze

Format: EVOLUTE® EXPRESS CX 10 mg Fixed Well Plate, part number 601-0010-PX01

Sample pre-treatment: Pipette patient urine (200 µL) into a container and add appropriate amount of internal standard. Pipette urine calibrators and QC samples (200 µL) and add appropriate amount of internal standard. Pre-treat each of the samples with HCl (0.01N, 200 µL) and gently vortex the solutions.

Sample loading: Load pre-treated samples onto wells. Allow to flow through sorbent via gravity.

Sample washing:

1) Add water (1000 µL) to each well and allow to flow by gravity.
2) Add 0.01N HCl/isopropanol (80:20, v/v, 1000 µL) to each well and allow to flow by gravity. Collect into waste container. Apply vacuum or positive pressure as needed between each step.

Analyte elution: Add a solution of tetrahydrofuran:methanol:acetonitrile:2% aqueous ammonium hydroxide (60:20:19:1, v/v/v/v, 600 µL) and allow solvent to gravity flow into a sample collection plate. Apply positive pressure or pull slight vacuum as needed during collection process to facilitate a flow rate of 1 mL per minute.

Post extraction: Add water (600 µL) to the eluent in each well. Load the collection plate onto the HPLC autosampler.

Additional information: Prepare elution solution at a 50 mL total volume by mixing THF (30 mL): MeOH (10 mL): Acetonitrile (9.5 mL): 2% aq NH4OH (0.5 mL). Mix thoroughly before use.

HPLC conditions

Instrument: Agilent 1200 Liquid Handling System (Agilent Technologies, Berkshire, UK)

Column: Phenomenex Gemini C18, 150 mm x 4.6 mm (5 µm)

Mobile phase: Solvent A: 5mM Ammonium Formate with 0.01% (v/v) Formic Acid
Solvent B: Acetonitrile with 0.01% (v/v) Formic Acid

Gradient:

Step

Time (min)

Flow Rate (µL/min)

%A

%B

1

0.0

1000

90

10

2

0.50

1000

90

10

3

2.5

1000

10

90

4

3.5

1000

10

90

5

4.0

1000

90

10

6

7.0

1000

90

10


Injection volume: 5 µL
Ion source temperature:  Ambient

Mass spectrometry conditions

Instrument: Applied Biosystems/MDS Sciex 4000 Q-Trap triple quadrupole mass
spectrometer (Applied Biosystems, Foster City, CA.) equipped with a Turbo Ionspray® interface for mass analysis.
Ion source temperature: 500 °C

Scan Function

Analyte

MRM

Transition

Declustering Potential (DP)

Collision Energy (CE)

Cell Exit Potential (CXP)

1

Alprozolam

308.8>280.5

30

35

16

2

Clonazepam

315.8>269.8

30

30

16

3

Diazepam

284.9>154

30

30

16

4

Flunitrazepam

313.9>267.9

30

30

16

5

Oxazepam

288>242

30

40

16

6

Temazepam

300.9>255

30

30

16

7

Nitrazepam

282.1>180

40

40

16

8

Normeperidine

234>160

30

30

16

9

Naltrexone

342.2>323.8

30

40

16

10

Morphine

286>165

45

40

16

11

Codeine

300>199

30

40

16

12

Oxymorphone

302>227

30

30

16

13

Oxycodone

316>241

45

40

16

14

Hydrocodone

300>199

25

25

16

15

6-Acetyl Codeine

342.4>255

30

40

16

16

6-Acetyl Morphine

328>165.5

30

45

16

17

Fentanyl

337>188

30

30

16

18

Buprenorphine

468.2>396.2

30

60

16

19

EDDP

278>234

30

40

16

Discussion

The advanced plate design technology utilized in the EVOLUTE® EXPRESS formats affords a LOAD-WASH-ELUTE strategy that yields a relatively abbreviated solid phase extraction methodology. Blank urine samples were fortified with basic drug standards to yield final sample concentration of 10 ng/mL. The fortified samples were diluted 1:1 with 0.01N HCL solution to pH adjust the urine samples and ionize the basic drugs in solution. These samples were then loaded onto the sorbent without the need for conditioning or equilibration steps. The sorbent was subsequently washed with water and then aqueous acidified isopropanol to remove urea, creatinine, salts, and other interferences.

The novel elution strategy utilizes tetrahydrofuran (THF) which is a water miscible organic solvent as the major polar elution solvent instead of ethyl acetate or other water immiscible organic solvents typically used to elute basic drugs. The THF can be mixed with other HPLC compatible solvents to yield an effective elution solvent that can be further diluted with water post elution and directly injected onto an HPLC cartridge. The novel solution of THF:MeOH:ACN:NH4OH (2%aq.) prepared as reported above, yields recoveries for the basic drug analytes ranging from 52–110% with typical averaged recoveries for most analytes greater than 75%. (Figure 2) The elution sample is collected as a clear solution prior to dilution and injection onto LC-MS/MS system. The percent relative standard deviation for each series was typically < 10%.

Figure 2. Plot of averaged recoveries for basic drugs extracted from fortified urine using EVOLUTE EXPRESS CX plates. The drug standards were fortified at 10 ng/mL.
 

Ordering information

Part number

Description

Quantity

601-0010-PX01  

 EVOLUTE® EXPRESS CX 10 mg Fixed Well Plate 

1

PPM-96   

Biotage® PRESSURE+ 96 Positive Pressure Manifold 96 well 

1

 

Literature number: AN806