Per- and polyfluoroalkyl substances (PFAS) continue to receive significant environmental and regulatory attention due to their persistence, bioaccumulation potential, and widespread occurrence in environmental and industrial matrices. The DIN EN 17892 standard provides a framework for determining selected PFAS compounds to support monitoring and compliance.
This application note presents a fully automated workflow for water samples using Biotage® EVOLUTE PFAS+ SPE cartridges, aligned with DIN EN 17892. The method, executed on the Biotage® PrepXpert-8 system with post-extraction concentration via TurboVap® LV, automates conditioning, loading, washing, and elution, minimizing manual handling and variability while enhancing reproducibility, throughput, and contamination control. In addition to the DIN EN 17892 targets, our research highlights additional PFAS compounds of relevance, expanding the scope for laboratories seeking consistent, traceable performance across a broader analyte list.
Table 1: Access here.
|
Operation |
Amount (mL) |
Input |
Output |
Flow (mL/min) |
|
Condition |
5 |
0.1% NH4OH in MeOH |
Solvent waste |
5 |
|
Condition |
5 |
MeOH |
Solvent waste |
5 |
|
Condition |
5 |
Water |
Water waste |
5 |
|
Load |
225 |
Sample |
Water waste |
5 |
|
Wash |
3 |
Water |
Water waste |
5 |
|
Purge |
3 |
Air |
Water waste |
10 |
|
Dry* |
20 min |
Nitrogen |
Water waste |
- |
|
Rinse |
6 |
MeOH |
Sample rinse |
60 |
|
Wait 15 sec |
||||
|
Elute |
10 |
Sample |
Vial B |
2 |
|
Elute |
5 |
0.1% NH4OH in MeOH |
Vial B |
2 |
|
Purge |
5 |
Nitrogen |
Vial B |
10 |
*N2 purge pressure for drying was increased from 1 bar to 1.5 bar to ensure complete drying of the adsorbent bed.
|
Bath temperature |
40 ˚C |
|
Evaporation mode |
Method (Ramp gradient) |
|
Manifold setup |
48 positions |
|
Rack row height |
120 mm* |
|
Step 1 |
1.2 L/min for 15 min |
|
Step 2 |
3.0 L/min for 15 min |
|
Step 3 |
3.5 L/min for 40 min |
*The nozzle position was adjusted such that it was as far to the right as possible to give the user a clear view of the vortex within the tube.
A: 20 mM ammonium acetate in water B: Methanol
Table below is the LC gradient
|
Time (min) |
%A |
%B |
|
0.00 |
90 |
10 |
|
0.50 |
90 |
10 |
|
2.00 |
60 |
40 |
|
7.50 |
0 |
100 |
|
9.00 |
0 |
100 |
|
9.10 |
90 |
10 |
|
10.00 |
90 |
10 |
For the work being done here, a total of five points were used in the calibration covering a range of 0.1-20 ppt. The curve was forced through zero and achieved excellent R2 values.
A target MRL of 1 ng/L was selected and at least seven replicate laboratory fortified blanks (LFBs) were created and ran at that concentration, with at least nine data points being selected for each compound. Figure 2 below illustrates the results of this test; the calculated detection limit (DL) for all DIN target compounds ranged from 0.09 ng/L (N-EtFOSAA) to 0.51 ng/L (PFTrDS) with 87% of the calculated DL being lower than 1/3 the concentration of the MRL. The total sum of the calculated DL for the select longer-chain target compounds such as PFOA, PFNA, PFHxS, and PFOS is less than 1 ng/L at 0.94 ng/L. The additional monitored compounds also performed reasonably well and achieved calculated DL ranging from 0.16 ng/L (8:2 FTUCA) to 0.60 ng/L (PFOcDA).
The data for individual compounds is shown in Appendix D.
The LFBs used to calculate the detection limit were also used to determine the precision and accuracy of the sample preparation process. Figure 3 illustrates the accuracy, while Figure 4 shows the precision. All target compounds, excluding two of the additional monitored compounds, recovered greater than 89% of the spiked amount and had calculated coefficient of variation (CV) less than 17%, which are in line with those observed in method DIN EN 17892. The additional monitored compounds FOSAA and PFOcDA were the only two targets that recovered less than 89% with recoveries of 50% and 80% respectively
at high calculated CV values. These compounds, as well as many of the later eluting compounds, have been observed to behave more inconsistently in the presence of residual water and consistency can be improved by increasing the column dry time or pressure if they are of particular concern.
A study was conducted to investigate each part of the extraction process and any their potential contribution to PFAS background. This sequential process is visualized below and highlights each component of the full extraction process as they are screened for PFAS background contribution before finally being combined into a single process when extracting the reagent water blank samples.
The results of these tests are given in Appendix F, and selected data are shown below in Figures 5-8.
For those results which were generated using only the analytical system, all target analytes were N.D. (unable to be separated from the noise in the baseline) and so were not listed out in the previous tables.
When examining the data resulting for all blank tests and the full laboratory reagent blank (LRB) tests (which includes the Biotage® PrepXpert-8, the EVOLUTE® PFAS+ SPE cartridges, and the TurboVap® LV) there are clear indications of the presence of a PFAS background.
However, even at the highest concentrations detected, all levels are much lower than the 1/3 MRL limit indicating that the background is acceptable and will not interfere with future sample runs. Specifically, PFBA can be seen in the background contribution from the EVOLUTE® PFAS+ SPE cartridges, however any background is flushed out during the conditioning steps of the extraction method, as indicated by the LRB resulting in non-detect levels of PFBA.
To simulate an influent sample, four LFB samples were created with concentrations which were more than two times greater than the highest point on the calibration curve. These samples were extracted, and the clean up procedure given in Appendix A was run three times. To ensure that the system background was adequately reduced, a set of four LRB samples were extracted immediately after the cleaning procedure and analysed. The LRB data obtained from this study is presented in Appendix G.
The trace levels of PFAS targets that were observed were non-detect for all compounds. These results show that the cleaning method given in Appendix A is sufficient to clean the PrepXpert-8 system; however, if higher than desired concentrations of any PFAS compounds remain, it is suggested to run additional cleaning methods to help re-establish the system background.
This study confirms that the Biotage® PrepXpert-8 system, combined with the TurboVap® LV, delivers a robust and efficient platform for PFAS extraction in accordance with DIN EN 17892. The system was verified to be PFAS-free, eliminating concerns of background contamination, and demonstrated full capability in extracting both the regulated compounds in DIN EN 17892 and an extended panel of additional PFAS targets.
By automating extraction and cleaning steps, the workflow reduces operator workload, enhances reproducibility, and supports scalable, traceable performance, offering a dependable solution for laboratories aiming to streamline PFAS sample preparation without compromising analytical quality.
Download appendix A to F here.
Literature number: AN1017