Hexane extractable material (HEM), also known as Oil & Grease extraction is a simple water quality measurement. The measurement can be used to assess pollution for compliance or as a measure of influent contamination to protect a waste treatment plant from levels that might cause malfunction.
This measurement is used around the world and regulatory agencies have developed methods such as US EPA 1664B, ISO 11349, and Standard Methods 5520G.1,2,3 We have shown the feasibility of Oil & Grease measurement on a dedicated automated system, the Biotage® Horizon 3100 (previously known as the SPE-DEX® 3100).4 In this work we will demonstrate the performance of a general purpose automated Oil & Grease and semi volatile extraction system, the Biotage® Horizon 5000.
The performance of the system for this application will be demonstrated by showing compliance with the quality control requirement of US EPA Method 1664B for small modifications. A new extraction procedure would be considered a modification to the method; therefore the laboratory is required to repeat the initial precision and recovery (IPR) study in Section 9.2.2 9 to demonstrate that the modification produces results equivalent to or superior to results produced by this method.
If the detection limit is suspected to be affected by the new modification a Method Detection Limit study (MDL) would need to be demonstrated. In this case the same disk and chemistry is used, only the automation has changed. 5 In this study the laboratory confirming the performance of each typical solid phase extraction (SPE) configuration on the Biotage® Horizon 5000 by conducting an IPR study to satisfy the quality control requirements for a method modification.
Table 1. Biotage® Horizon 5000 Extraction Method for 90 mm Pacific® Premium Disks (No Pre-Filters).
|
Step |
Solvent |
Solvent Vol. (mL) |
Purge Time (s) |
Pump Rate (s) |
Sat. Time (s) |
Soak Time (s) |
Drain Time (s) |
|
|
1. Condition SPE Disk |
Hexane |
16 |
60 |
2 |
1 |
60 |
60 |
|
|
2. Condition SPE Disk |
MeOH |
16 |
60 |
2 |
1 |
60 |
2 |
|
|
Step |
|
Sample Flow Rate (#) |
Done Loading Sample Delay (s) |
|
|
|
|
|
|
6. Load Sample |
|
5 |
45 |
|
|
|
|
|
|
Step |
|
Dry Time (s) |
Pump Rate (#) |
N2 Blanket |
|
|
|
|
|
7. Air Dry Disk Timer |
|
180 |
6 |
Off |
|
|
|
|
|
Step |
Solvent |
Solvent Vol. (mL) |
Purge Time (s) |
Pump Rate (#) |
N2 Blanket |
Sat. Time (s) |
Soak Time (s) |
Elute Time (s) |
|
8. Elute Sample Container 9. Elute Sample Container |
Hexane Hexane |
28 22 |
35 35 |
5 5 |
Off Off |
1 1 |
20 45 |
30 45 |
|
10. Elute Sample Container |
Hexane |
22 |
60 |
6 |
Off |
1 |
45 |
60 |
|
Step |
Solvent |
Solvent Vol. |
Purge |
Pump |
N2 Blanket |
Sat. Time (s) |
Soak Time (s) |
Drain Time (s) |
|
11. Wash Sample Container |
MeOH |
16 |
60 |
6 |
Off |
1 |
45 |
60 |
|
Step |
Solvent |
Solvent Vol. |
Purge |
Pump |
N2 Blanket |
Sat. Time (s) |
Soak Time (s) |
Elute Time (s) |
|
12. Elute Sample Container |
Hexane |
16 |
35 |
5 |
Off |
1 |
45 |
45 |
|
13. Elute Sample Container |
Hexane |
16 |
35 |
5 |
Off |
1 |
45 |
45 |
|
14. Elute Sample Container |
Hexane |
16 |
60 |
6 |
Off |
1 |
45 |
60 |
Table 2. Biotage® Horizon 5000 Extraction Method for 47 mm Pacific® Premium Disks (No Pre-Filters).
|
Step |
Solvent |
Solvent Vol. (mL) |
Purge Time (s) |
Pump Rate (#) |
Sat. Time (s) |
Soak Time (s) |
Drain Time (s) |
Sample Flow Rate (#) |
Done Loading Sample Delay (s) |
Dry Time (s) |
N₂ Blanket |
Elute Time (s) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
1. Condition SPE Disk |
Hexane |
11 |
60 |
2 |
1 |
60 |
60 |
|||||
|
2. Condition SPE Disk |
MeOH |
11 |
60 |
2 |
1 |
60 |
2 |
|||||
|
6. Load Sample |
5 |
45 |
||||||||||
|
7. Air Dry Disk Timer |
6 |
180 |
Off |
|||||||||
|
8. Elute Sample Container |
Hexane |
16 |
35 |
5 |
1 |
20 |
Off |
30 |
||||
|
9. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
10. Elute Sample Container |
Hexane |
11 |
60 |
6 |
1 |
45 |
Off |
60 |
||||
|
11. Wash Sample Container |
MeOH |
11 |
60 |
6 |
1 |
45 |
60 |
Off |
||||
|
12. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
13. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
45 |
Off |
45 |
Table 3. Biotage® Horizon 5000 Extraction Method for 90 mm Pacific® Premium Disks with Pre-Filters.
|
Step |
Solvent |
Solvent Vol. (mL) |
Purge Time (s) |
Pump Rate (#) |
Sat. Time (s) |
Soak Time (s) |
Drain Time (s) |
Sample Flow Rate (#) |
Done Loading Sample Delay (s) |
Dry Time (s) |
N₂ Blanket |
Elute Time (s) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
1. Condition SPE Disk |
Hexane |
16 |
60 |
2 |
1 |
60 |
60 |
|||||
|
2. Condition SPE Disk |
MeOH |
16 |
60 |
2 |
1 |
60 |
2 |
|||||
|
6. Load Sample |
5 |
45 |
||||||||||
|
7. Air Dry Disk Timer |
6 |
480 |
Off |
|||||||||
|
8. Elute Sample Container |
Hexane |
28 |
35 |
5 |
1 |
25 |
Off |
30 |
||||
|
9. Elute Sample Container |
Hexane |
22 |
35 |
5 |
1 |
50 |
Off |
45 |
||||
|
10. Elute Sample Container |
Hexane |
22 |
60 |
6 |
1 |
50 |
Off |
60 |
||||
|
11. Wash Sample Container |
MeOH |
16 |
60 |
6 |
1 |
50 |
60 |
Off |
||||
|
12. Elute Sample Container |
Hexane |
16 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
13. Elute Sample Container |
Hexane |
16 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
14. Elute Sample Container |
Hexane |
16 |
60 |
6 |
1 |
45 |
Off |
60 |
Table 4. Biotage® Horizon 5000 Extraction Method for 47 mm Pacific® Premium Disks with Pre-Filters.
|
Step |
Solvent |
Solvent Vol. (mL) |
Purge Time (s) |
Pump Rate (#) |
Sat. Time (s) |
Soak Time (s) |
Drain Time (s) |
Sample Flow Rate (#) |
Done Loading Sample Delay (s) |
Dry Time (s) |
N₂ Blanket |
Elute Time (s) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
1. Condition SPE Disk |
Hexane |
11 |
60 |
2 |
1 |
60 |
60 |
|||||
|
2. Condition SPE Disk |
MeOH |
11 |
60 |
2 |
1 |
60 |
2 |
|||||
|
6. Load Sample |
5 |
45 |
||||||||||
|
7. Air Dry Disk Timer |
6 |
300 |
Off |
|||||||||
|
8. Elute Sample Container |
Hexane |
16 |
35 |
5 |
1 |
23 |
Off |
30 |
||||
|
9. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
48 |
Off |
45 |
||||
|
10. Elute Sample Container |
Hexane |
11 |
60 |
6 |
1 |
48 |
Off |
60 |
||||
|
11. Wash Sample Container |
MeOH |
11 |
60 |
6 |
1 |
48 |
60 |
Off |
||||
|
12. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
13. Elute Sample Container |
Hexane |
11 |
35 |
5 |
1 |
45 |
Off |
45 |
||||
|
14. Elute Sample Container |
Hexane |
11 |
60 |
6 |
1 |
45 |
Off |
60 |
Table below shows the initial demonstration of precision and recovery for the 90 mm disks without pre-filters.
|
Sample |
Recovery (mg) |
Recovery (%) |
|---|---|---|
|
1 |
39.2 |
98.0 |
|
2 |
38.0 |
95.0 |
|
3 |
38.0 |
95.0 |
|
4 |
38.4 |
96.0 |
|
Average Recovery (%): |
96.0 |
|
|
Standard Deviation (%): |
1.41 |
|
Table below shows the same information for 90 mm disks with pre-filters
|
Sample |
Recovery (mg) |
Recovery (%) |
|---|---|---|
|
1 |
38.2 |
95.5 |
|
2 |
38.1 |
95.3 |
|
3 |
38.0 |
95.0 |
|
4 |
38.7 |
96.7 |
|
Average Recovery (%): |
95.6 |
|
|
Standard Deviation (%): |
0.746 |
|
Table below shows the results for 4 samples using 47 mm disks without pre-filters.
|
Sample |
Recovery (mg) |
Recovery (%) |
|---|---|---|
|
1 |
39.3 |
98.3 |
|
2 |
39.2 |
98.0 |
|
3 |
39.4 |
98.5 |
|
4 |
39.7 |
99.2 |
|
Average Recovery (%): |
98.5 |
|
|
Standard Deviation (%): |
0.510 |
|
Table below shows the results for 4 samples using 47 mm disks with pre-filters
|
Sample |
Recovery (mg) |
Recovery (%) |
|---|---|---|
|
1 |
39.6 |
99.0 |
|
2 |
39.1 |
97.8 |
|
3 |
38.4 |
96.0 |
|
4 |
38.8 |
97.0 |
|
Average Recovery (%): |
97.5 |
|
|
Standard Deviation (%): |
1.27 |
|
In each case the recovery is better than 95%, well within the acceptable range specified in method 1664B of 83–101%. The standard deviation is well below the 11% limit specified in the method for each of the four configurations tested.
Literature number: AN118-HOR