Figure 1: Structure of melamine
This application note describes the extraction of melamine from infant formula using EVOLUTE® EXPRESS CX polymeric mixed-mode strong cation exchange SPE cartridges.
Melamine is traditionally used in making plastics, however, it’s low cost and high nitrogen content has led to exploitation in various sections of the food industry, most notably involving dairy products. The standard test for estimating protein content is based on measurement of nitrogen levels, therefore, addition of melamine to sub standard or watered down milk results in the protein levels appearing higher. Sustained melamine exposure can result in kidney stones and renal failure, with the young being most susceptible.
EVOLUTE® EXPRESS CX SPE cartridges and plates extract a wide range of basic analytes from biological fluids and other aqueous matrices. EVOLUTE® EXPRESS CX removes matrix components such as proteins, salts, non-ionizable interferences and phospholipids, delivering cleaner extracts with reproducible recoveries for accurate quantitation.
Melamine
EVOLUTE® EXPRESS CX 60 mg/ 3 mL (Tables) SPE cartridges, part number 611-0006-BXG.
[Note: This application was originally developed using EVOLUTE® CX 50 mg/3 mL SPE cartridges, p/n 611-0005-B. This format has now been replaced by the EXPRESS format, which provides improved flow characteristics].
For powdered formula: make up as per manufacturer’s instructions, and cool prior to extraction. Dilute all milk samples (1 mL) with ammonium acetate buffer (1:1, v/v, 50 mM, pH 5).
Condition cartridge with methanol (3 mL).
Equilibrate cartridge with ammonium acetate buffer (50 mM, pH 5, 3 mL).
Load pre-treated sample (2 mL) at a flow rate of 3 mL /min.
Remove polar and ionic interferences with ammonium acetate buffer (50 mM, pH 5, 3 mL).
Remove non-polar interferences with methanol (3 mL).
Elute melamine with methanol containing ammonium hydroxide (95:5, v/v, 3 mL).
Evaporate to dryness at 40°C and reconstitute in acetonitrile/ water (90:10, v/v, 500 μL) prior to analysis.
Waters 2795 Liquid Handling System
Phenomenex Luna HILIC 3 μm analytical cartridge (100 x 2.0 mm id)
Phenomenex Luna Phenyl-Hexyl security guard cartridge
Isocratic, acetonitrile:20mM Ammonium Formate pH 3.2 (75:25, v/v), flow rate of 0.3 mL/min.
20 μL
Ambient
Waters Ultima Pt triple quadrupole mass spectrometer equipped with an electrospray interface
350 °C
100 °C
2.4 x 10-³ mbar
Positive ions were acquired in the multiple reaction monitoring mode (MRM).
127 >85 (collision energy 12 eV)
127 >68 (collision energy 14 eV)
From the results it can be seen that recoveries from a range of infant formula products were above 85% with <10% relative standard deviations below 10.
|
Matrix |
Powdered Baby Milk |
Liquid Baby Milk 1 |
Liquid Baby Milk 2 |
Liquid Baby Milk 3 |
|
Melamine %recovery |
84 |
97 |
89 |
98 |
|
% RSD (n=6) |
<10 |
<10 |
<10 |
<10 |
This application note is based on the poster ‘Extraction of melamine from various matrices using resin based mixed-mode cation exchange SPE and analysis with LC-MS/MS’, L Williams et al, presented at ASMS, Philadelphia, May 31st June 4th, 2009.
|
Part number |
Description |
Quantity |
|
611-0006-BXG |
EVOLUTE® EXPRESS CX 60 mg/3 mL (Tables) SPE Cartridges |
50 |
Literature number: AN732