Figure 1. Structure of Δ9-tetrahydrocannabinol (THC).
This application note describes a procedure for sample pre-treatment and extraction of THC and metabolites from human hair, using Biotage® Lysera for matrix pulverization of the sample prior to clean up using ISOLUTE® SLE+ supported liquid extraction.
Manual processing protocols were developed using the Biotage® PRESSURE+ 96 (plate format) or 48 (column format) positive pressure manifolds. For automated processing, protocols were developed using Biotage® ExtraheraTM .
The application note contains procedures optimized for both individual column format and 96-well plate format for higher throughput applications. The methodology delivers clean extracts with analyte recoveries >75% (plate format) or >60% (column format) with %RSD <10% for most analytes and LLOQ from 200 fg/mg of hair.
Both manual and automated procedures gave comparable results.
ISOLUTE® SLE+ Supported Liquid Extraction cartridges and plates offer an efficient alternative to traditional liquid- liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation and significantly reduced preparation time.
Tetrahydrocannabinol (THC), 11-Nor-9-carboxy-Δ⁹- tetrahydrocannabinol (THC-COOH), 11-Hydroxy-Δ⁹- tetrahydrocannabinol (THC-OH), Δ9-tetrahydrocannabinolic acid-A (THCAA), cannabidiol (CBD), and cannabinol (CBN)
Tetrahydrocannabinol –D3 (THC- D3), 11-Nor-9-carboxy-Δ⁹- tetrahydrocannabinol-D3 (THC- COOH-D3) and 11-Hydroxy-Δ⁹- tetrahydrocannabinol- D3 (THC-OH- D3)
ISOLUTE® SLE+ 400 µL capacity cartridges (p/n 820-0055-B) or ISOLUTE® SLE+ 400 µL capacity plates (p/n 820-0400-P01)
Weigh 20 mg of hair into 2 mL Biotage® Lysera tubes containing 4 x 2.4 mm stainless steel beads. Add 1 mL methanol to each hair sample. Also add 10 µL of a 100 pg/mL internal standard solution making a 50 pg/mg spike.
Grind to a fine powder using Biotage® Lysera: 3 x 5.3 m/sec for 3 minutes with a 20 sec dwell.
Centrifuge tubes for 10 minutes at 13,300 rpm (Heraeus Pico 17 Microcentrifuge (Thermo Scientific) with 24 position, 2 mL rotor).
Transfer 200 µL of supernatant into 12 x 75 mm glass tubes or 2 mL collection plates and evaporate extracts using a TurboVap® LV at 20 °C or Biotage®SPE Dry 96 depending on the format being used.
Reconstitute in methanol:water (70:30, v/v. 200 µL).
|
|
ISOLUTE® SLE+ 400 µL Cartridges |
ISOLUTE® SLE+ 400 µL Plate |
|
Sample Loading |
Load 200 µL of reconstituted extract onto the ISOLUTE® SLE+ cartridge. A pulse of pressure is not needed as the methanolic extract flows straight onto the bed. Allow the sample to absorb for 5 minutes. |
Load 200 µL of reconstituted extract onto the ISOLUTE® SLE+ well. A pulse of pressure is not needed as the methanolic extract flows straight onto the bed. Allow the sample to absorb for 5 minutes. |
|
Analyte Extraction |
Apply MTBE (600 µL) allow to flow under gravity for 5 minutes. Apply a further aliquot of MTBE (600 µL) and allow to flow under gravity for 5 minutes. For complete removal apply a pulse of positive pressure at 10 psi (10–20 seconds). |
Apply MTBE (600 µL) allow to flow under gravity for 5 minutes. Apply a further aliquot of MTBE (600 µL) and allow to flow under gravity for 5 minutes. For complete removal apply a pulse of positive pressure at 10 psi (10–20 seconds). |
|
Collection Vessels |
Collect extract in 12 x 75 mm glass tubes. |
Collect extract in 96-well collection plates. |
|
Post Elution |
Evaporate extracts to dryness at 40 °C, for 30 minutes at a flow rate of 1.5 L/min using a TurboVap® LV. |
Evaporate extracts to dryness at 40 °C, for 30 minutes at a flow rate of 20–40 L/min using a Biotage® SPE Dry 96. |
|
Reconstitute |
Reconstitute extracts in a mix of mobile phase A/mobile phase B (70:30, v/v, 200 µL). |
Reconstitute extracts in a mix of mobile phase A/mobile phase B (80:20, v/v, 200 µL). Vortex mix. |
|
|
Vortex mix and transfer to a 96-well format plate and cover with a sealing mat prior to injection. |
Cover plate with a sealing mat prior to injection. |
Shimadzu Nexera X2 UHPLC
ACE Excel 2 C18 (50 x 2.1 mm) with a Restek EXP holder and Restek C18 guard column
A: 0.01% Acetic Acid (aq)
B: 0.01% Acetic Acid in MeOH
0.3 mL/min
50 °C
5 µL
Table 1. UHPLC gradient.
|
Time (min) |
%A |
%B |
|
0 |
50 |
50 |
|
0.5 |
20 |
80 |
|
2.00 |
10 |
90 |
|
4.00 |
10 |
90 |
|
4.01 |
50 |
50 |
Shimadzu 8060 Triple Quadrupole MS using ES interface
3 L/min
5 L/min
15 L/min
400 oC
300 oC
500 oC
270 kPa
Table 2. MS conditions for target analytes in positive and negative mode.
|
Analytes |
MRM Transition |
Collision Energy |
|
THC-D3 |
318.0 > 196.15 |
-24 |
|
318.0 > 123.2 |
-32 |
|
|
THC |
315.0 > 193.10 |
-23 |
|
315.0 > 123.2 |
-32 |
|
|
OH-THC-D3 |
334.0 > 316.15 |
-15 |
|
334.0 > 196.25 |
-25 |
|
|
OH-THC |
331.0 > 313.3 |
-15 |
|
331.0 > 193.25 |
-26 |
|
|
THC-COOH-D3 |
346.3 > 302.3 |
22 |
|
346.3 > 248.30 |
28 |
|
|
THC-COOH |
343.3 > 299.3 |
22 |
|
343.3 > 245.25 |
30 |
|
|
CBN |
311.0 > 223.0 |
-22 |
|
311.0 > 241.2 |
-17 |
|
|
CBD |
313.2 > 245.15 |
24 |
|
313.2 > 179.25 |
20 |
This simple sample preparation method delivers clean extracts and analyte recoveries mostly greater than 75% with RSDs lower than 10% for all analytes (see fig 2), and LLOQs below 10 pg/mg and as low as 200 fg/mg for THC-COOH and (see table 3) for all ISOLUTE® SLE+ formats used.
Figure 2. Average analyte recoveries and %RSD (n=7) for ISOLUTE® SLE+ cartridge and plate formats.
Calibration curve performance was investigated from hair samples spiked between 0.1–200 pg/mg of hair. Good linearity was observed for all analytes typically delivering r2 values greater than 0.99. Table 3. details linearity performance and associated LOQ for each analyte using the ISOLUTE® SLE+ cartridge format. Similar results were achieved using the 96-well plate format.
Table 3. Analyte calibration curve r2 and LOQ performance.
|
|
Column Format |
Plate Format |
||
|
Analytes |
r2 |
LLOQ (pg/mg) |
r2 |
LLOQ (pg/mg) |
|
THC |
0.997 |
10 |
0.998 |
10 |
|
OH-THC |
0.997 |
10 |
0.998 |
10 |
|
THC-COOH |
0.997 |
0.2 |
0.997 |
0.2 |
|
CBN |
0.997 |
10 |
0.997 |
10 |
|
CBD |
0.997 |
1 |
0.995 |
0.5 |
Figure 3. Calibration curves for THC (a), OH-THC (b), THC-COOH (c), CBD (d) and CBN (e) using human hair with 400 µL capacity column format (loading 200 µL of extracted sample as described).
|
Part Number |
Description |
Quantity |
|
19-060 |
Biotage® Lysera |
1 |
|
19-649 |
2 mL Reinforced Tubes with screw caps (Bulk pack) |
1000 |
|
19-640 |
2.4 mm Metal Beads - 500 grams |
1 |
|
820-0055-B |
ISOLUTE® SLE+ 400 µL Sample Volume Cartridges |
50 |
|
820-0400-P01 |
ISOLUTE® SLE+ 400 µL Capacity Plate |
1 |
|
PPM-96 |
Biotage® PRESSURE+ 96 Positive Pressure Manifold |
1 |
|
PPM-48 |
Biotage® PRESSURE+ 48 Positive Pressure Manifold |
1 |
|
415000 |
TurboVap® LV |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry 96 Sample Evaporator 220/240 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry 96 Sample Evaporator 100/120 V |
1 |
|
121-5203 |
Collection Plate, 2 mL Square |
50 |
|
121-5204 |
Piercable Sealing Mat |
50 |
|
C44651 |
Test Tubes (12 x 75 mm, Uncapped) |
1000 |
|
414001 |
Biotage® Extrahera™ |
1 |
The method described in this application note was automated on the Biotage® Extrahera™ using ISOLUTE® SLE+ 400 µL capacity cartridges and 96-well plates. This appendix contains the software settings required to configure Extrahera to run the column format method. As described in the main body of the application note, analyte recoveries, %RSDs, linearities and LOQs were comparable for both manually processed and automated methods, for both extraction formats.
|
Sample Name: |
THC Hair – Pre concentration Method |
|
Sample Plate/Rack: |
12 x 75 mm Test Tubes, 24 |
|
Extraction Media: |
ISOLUTE® SLE+ 400 µL Cartridges |
|
Solvent Description |
|
|
1 |
MTBE |
|
2 |
|
|
3 |
|
|
4 |
|
|
5 |
|
|
6 |
|
|
7 |
|
|
8 |
|
|
9 |
|
|
10 |
|
Solvent |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
10 |
|
Reservoir Type |
Refillable |
Non Refillable |
||||||||
|
Capacity |
||||||||||
|
Aspiration flow rate (mL/min) |
10 |
|||||||||
|
Dispense flow rate (mL/min) |
10 |
|||||||||
|
Lower air gap flow rate (mL/min) |
10 |
|||||||||
|
Lower air gap volume (µL) |
5 |
|||||||||
|
Upper air gap flow rate (mL/min) |
120 |
|||||||||
|
Upper air gap volume (µL) |
100 |
|||||||||
|
Upper air gap dispense pause |
300 |
|||||||||
|
Conditioning? |
Yes |
|||||||||
|
Conditioning number of times |
2 |
|||||||||
|
Conditioning flow rate (mL/min) |
10 |
|||||||||
|
Chlorinated |
No |
|||||||||
|
Serial dispense |
No |
|||||||||
Literature Number: AN919