Using a semi-preparative normal phase HPLC column on Biotage® Selekt

Written by Biotage | Nov 23, 2026, 8:24:44 AM

The Biotage® Selekt chromatography instrument can be run at a maximum flowrate of 300 mL/min or a maximum pressure of 30 bar. These high flowrates and pressures enables a user to do chromatography using not only dry-packed, single-use plastic flash columns containing small (≥20 μm) spherical silica  particles, but also semi-preparative, slurry-packed  HPLC columns for multiple use with smaller (≤20 μm) spherical silica particles.

In this application note, we show the separation of a three component mixture using a semi-preparative slurry-packed HPLC column with 15 μm spherical silica particles and a Biotage® Selekt instrument.

Experimental

A sample of equal amounts of three similar compounds,  dimethyl phthalate, diethyl phthalate, and dibutyl phthalate (Figure 1), was separated using a commercially available 15 μm Silica 250 x 21.2 mm column on a Biotage® Selekt instrument.

Figure 1. Compounds separated. 1) Dimethyl phthalate 2) Diethyl phthalate 3) Dibutyl phthalate. 

Sample solution

  • Dimethyl phthalate (1): 5.0 g

  • Diethyl phthalate (2): 5.0 g

  • Dibutyl phthalate (3) 5.0 g

  • n-Heptane: 10 mL

  • Ethyl acetate: 2 mL

The material was dissolved in 10 mL n-heptane in a 25 mL measurement flask and the solution was made up to 25.0 mL with 2 mL ethyl acetate. This gave a total concentration of 0.60 g/mL.

The semi-preparative HPLC column (slurry-packed in ethanol)  was attached to the Biotage® Selekt instrument equipped with a 5 mL sample loop with an inner diameter of 1.6 mm and a manual injection valve using PTFE tubing with 1.6 mm inner diameter (Figure 2).

Figure 2. The semi-preparative HPLC column setup on Biotage® Selekt.

A new column, with the following characteristics, was created in the "Column Administration" section of the Biotage® Selekt software (Figure 3):

  • Name: 15 μm Silica 250 x 21 mm

  • Column volume: 58 mL

  • Equilibration flow rate: 120 mL/min

  • Equilibration pressure: 30 bar

  • Equilibration length: 3 CV

  • Default flow rate: 120 mL/min

  • Max pressure: 30 bar

Figure 3. Column edit screen on Biotage® Selekt.

The HPLC column was washed twice with 80 mL n-heptane– ethyl acetate 1:1 at 100 mL/min (this gave a back pressure of approx. 22 bar), and then equilibrated with 3 CV n-heptane– ethyl acetate 85:15 at 120 mL/min (also approx. 22 bar). 5.0 mL sample (1.80 g, approx. 3.6% loading) was added via the sample loop and the separation was executed using the method below.

Chromatography conditions

  • Solvent A: n-Heptane, Solvent B: Ethyl acetate

  • Column equilibration: 15% B, 3 CV, 120 mL/min

  • Gradient run: 15% B for 1 CV, 15–50% B for 4 CV, 50% B for 1 CV

    UV1:254 nm (Monitor), UV2: 280 nm (Monitor),

    λ-All: 200–400 nm (Collect), baseline correction: On, Threshold: 200 mAU

  • Column: 15 μm Silica 250 x 21.2 mm

  • Flow rate: 120 mL/min

Results

The separation worked smoothly and the pressure did not exceed 27 bar during the chromatography. The compounds were separated with excellent baseline separation (Figure 4).

Figure 4. Chromatogram of the separation.

Conclusions

In this application note, we show that a semi-preparative slurry-packed HPLC column with 15 μm spherical silica particles can easily be used on a Biotage® Selekt instrument.  The separation of a three-component mixture using this setup worked smoothly.

Literature number: AN138