Using a semi-preparative normal phase HPLC column on Biotage® Selekt

The Biotage® Selekt chromatography instrument can be run at a maximum flowrate of 300 mL/min or a maximum pressure of 30 bar. These high flowrates and pressures enables a user to do chromatography using not only dry-packed, single-use plastic flash columns containing small (≥20 μm) spherical silica  particles, but also semi-preparative, slurry-packed  HPLC columns for multiple use with smaller (≤20 μm) spherical silica particles.

In this application note, we show the separation of a three component mixture using a semi-preparative slurry-packed HPLC column with 15 μm spherical silica particles and a Biotage® Selekt instrument.

Experimental

A sample of equal amounts of three similar compounds,  dimethyl phthalate, diethyl phthalate, and dibutyl phthalate (Figure 1), was separated using a commercially available 15 μm Silica 250 x 21.2 mm column on a Biotage® Selekt instrument.

AN138 - Figure 1 - Compounds separated

Figure 1. Compounds separated. 1) Dimethyl phthalate 2) Diethyl phthalate 3) Dibutyl phthalate. 

Sample solution

  • Dimethyl phthalate (1): 5.0 g

  • Diethyl phthalate (2): 5.0 g

  • Dibutyl phthalate (3) 5.0 g

  • n-Heptane: 10 mL

  • Ethyl acetate: 2 mL

The material was dissolved in 10 mL n-heptane in a 25 mL measurement flask and the solution was made up to 25.0 mL with 2 mL ethyl acetate. This gave a total concentration of 0.60 g/mL.

The semi-preparative HPLC column (slurry-packed in ethanol)  was attached to the Biotage® Selekt instrument equipped with a 5 mL sample loop with an inner diameter of 1.6 mm and a manual injection valve using PTFE tubing with 1.6 mm inner diameter (Figure 2).

AN138 - Figure 2 - HPLC setup on Selekt

Figure 2. The semi-preparative HPLC column setup on Biotage® Selekt.

A new column, with the following characteristics, was created in the "Column Administration" section of the Biotage® Selekt software (Figure 3):

  • Name: 15 μm Silica 250 x 21 mm

  • Column volume: 58 mL

  • Equilibration flow rate: 120 mL/min

  • Equilibration pressure: 30 bar

  • Equilibration length: 3 CV

  • Default flow rate: 120 mL/min

  • Max pressure: 30 bar

AN138 - Figure 3 - Column edit screen

Figure 3. Column edit screen on Biotage® Selekt.

The HPLC column was washed twice with 80 mL n-heptane– ethyl acetate 1:1 at 100 mL/min (this gave a back pressure of approx. 22 bar), and then equilibrated with 3 CV n-heptane– ethyl acetate 85:15 at 120 mL/min (also approx. 22 bar). 5.0 mL sample (1.80 g, approx. 3.6% loading) was added via the sample loop and the separation was executed using the method below.

Chromatography conditions

  • Solvent A: n-Heptane, Solvent B: Ethyl acetate

  • Column equilibration: 15% B, 3 CV, 120 mL/min

  • Gradient run: 15% B for 1 CV, 15–50% B for 4 CV, 50% B for 1 CV

    UV1:254 nm (Monitor), UV2: 280 nm (Monitor),

    λ-All: 200–400 nm (Collect), baseline correction: On, Threshold: 200 mAU

  • Column: 15 μm Silica 250 x 21.2 mm

  • Flow rate: 120 mL/min

Results

The separation worked smoothly and the pressure did not exceed 27 bar during the chromatography. The compounds were separated with excellent baseline separation (Figure 4).

AN138 - Figure 4 - Chromatogram of separation

Figure 4. Chromatogram of the separation.

Conclusions

In this application note, we show that a semi-preparative slurry-packed HPLC column with 15 μm spherical silica particles can easily be used on a Biotage® Selekt instrument.  The separation of a three-component mixture using this setup worked smoothly.

Literature number: AN138

 