Automated PFAS sample preparation for ISO 21675

Written by Biotage | Nov 23, 2026, 7:50:20 AM

Introduction

Increased awareness of PFAS contamination has driven the need for reliable analytical methods to support environmental monitoring initiatives. Due to their extensive historical use and environmental persistence, PFAS compounds are now routinely targeted in water quality assessments throughout the world. ISO 21675 provides a standardized protocol for the extraction and determination of selected PFAS compounds in water, promoting accuracy and comparability of results between laboratories. In this work, an automated workflow employing the Biotage® PrepXpert-8 system and EVOLUTE® PFAS cartridges was evaluated for compliance with ISO 21675 sample preparation and performance requirements. Following extraction, samples were concentrated using the TurboVap® LV platform, resulting in a streamlined and reproducible process well suited for high-throughput PFAS testing laboratories.

Analytes

 

Table 1. Listing of ISO 21675 target analytes

Analyte name

Acronym

CAS

ISO 21675 target compounds

   

Perfluoro-n-butanoic acid

PFBA

375–22–4

Perfluoro-n-pentanoic acid 

PFPeA 

2706–90–3

Perfluoro-n-butanesulfonic acid 

PFBS 

375–73–5

Perfluoro-n-hexanoic acid

PFHxA 

307–24–4

Hexafluoropropylene oxide dimer acid

HFPO-DA (GenX)

13252–13–6

Perfluoro-n heptanoic acid 

PFHpA 

375–85–9

Perfluoro-n hexanesulfonic acid 

PFHxS  

355–46–4

4,8-Dioxa-3-H-perfluorononanoic acid

ADONA (DONA)

919005–14–4

Perfluoro-n heptanesulfonic acid

PFHpS

375–92–8

6:2 Fluorotelomer sulfonic acid

6:2 FTS

27619–97–2

Perfluoro-n-octanoic acid 

PFOA 

335–67–1

Perfluoro-n-octanesulfonic acid

PFOS 

1763–23–1

Perfluoro-n-nonanoic acid 

PFNA  

375–95–1

8:2 Fluorotelomer unsaturated carboxylic acid

8:2 FTUCA

70887-84-2

Perfluoro-n decanoic acid

PFDA 

335–76–2

9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid

9Cl-PF3ONS

73606–19–6

8:2 Fluorotelomer sulfonic acid

8:2 FTS

39108–34–4

N-methyl perfluorooctanesulfonamidoacetic acid

N-MeFOSAA

2355-31-9

Perfluorooctanesulfonamide

FOSA

754–91–6

Perfluoro-n-decanesulfonic acid 

PFDS  

335–77–3

Perfluoro-n-undecanoic acid

 PFUnA (PFUnDA)

2058–94–8

N-ethyl perfluorooctanesulfonamidoacetic acid

N-EtFOSAA

2991-50-6

Perfluoro-n-dodecanoic acid 

PFDoA (PFDoDA)

307–55–1

Perfluoro-n-tridecanoic acid 

PFTrDA 

72629–94–8

N-methyl perfluorooctanesulfonamide

N-MeFOSA

31506-32-8

N-ethyl perfluorooctanesulfonamide

N-EtFOSA

4151-50-2

Perfluoro-n-tetradecanoic acid

PFTeDA

376-06-7

Perfluoro-n-hexadecanoic acid

PFHxDA

67905-19-5

8:2 Polyfluoroalkyl phosphate diester

8:2 diPAP

678-41-1

Perfluoro-n-octadecanoic acid

PFOcDA

16517-11-6

Solution preparation

Ammonia methanol solution

  1. Add 400 μL of NH4OH to every 99.6 mL of methanol in a clean beaker.
  2. Mix thoroughly to homogenize the solution.
  3. Prepare fresh solution daily before extraction.

Sample preparation procedure

  1. Clean the Biotage® PrepXpert-8 automated SPE system using the procedure given in appendix A. If the cleanliness of a particular channel is in question, it is recommended to repeat the cleaning procedure and verify channel cleanliness by extracting blanks.
  2. Set up and fill new sample containers with water.
  3. Adjust the sample pH to 3 using glacial acetic acid.
  4. Spike the extracted labeled standards into each sample at a concentration corresponding to the midpoint of the calibration curve.
  5. If necessary to perform statistical studies, spike a known amount of PFAS target standard into the water samples to achieve the desired concentration.
  6. Attach each prepared water sample container to the sample bottle cap on the Biotage® PrepXpert-8 system and ensure that the sip tube is positioned to drain all liquid from the sample bottle.
  7. Attach the EVOLUTE® PFAS SPE cartridges to the adapters of the Biotage® PrepXpert-8 system.
  8. Load 15 mL centrifuge tubes onto the collection vessel rack and ensure that the rack is installed on the Biotage® PrepXpert-8 automated tray.
  9. Program in the parameters for the extraction of samples by ISO 21675 given in table 2, adjust the sample-load volume as required, and run the method. The protocol will result in an approximately 10 mL extract.
  10. Determine the initial sample volume by either measuring the volume required to refill the sample container to the original mark using a graduated cylinder or by weighing the container before and after sample loading.
  11. Transfer the centrifuge tubes to the TurboVap® LV system and concentrate the extract to a volume of less than 0.5 mL using nitrogen according to the parameters in Table 3.
  12. Add methanol to each extract to achieve a final volume of 0.5 – 1.0 mL.
  13. Add an appropriate amount of internal standard to each extract and mix to homogenize.
  14. Inject the extract into a calibrated LC-MS/MS system and analyze it using the conditions given in the following sections.

Table 2. Extraction parameters

Operation

Amount (mL or time)

Input

Output

Flow (mL/min)

Condition

5

0.1% NH4OH in MeOH

Solvent waste

5

Condition

5

MeOH

Solvent waste

5

Condition

5

Water

Water waste

5

Load

265

Sample

Water waste

10

Wash

5

Water

Water waste

5

Purge

5

Air

Water waste

5

Dry

5 min

Nitrogen

Water waste

-

Rinse

5

0.1% NH4OH in MeOH

Sample

60

Wait

15 sec

Elute

8

Sample

Vial 2

5

Rinse

5

0.1% NH4OH in MeOH

Sample

60

Wait

15 sec

Elute

8

Sample

Vial 2

5

Purge

5

Air

Vial 2

5

 

Table 3. TurboVap® LV concentration protocol

Bath temperature:

60 °C

Evaporation mode:

Method (ramp gradient)

Manifold setup:

48 positions

Rack row height:

120 mm

Step 1:

1.5 L/min for 20 min

Step 2:

3.0 L/min for 15 min

Step 3:

3.5 L/min for 50 min

 

 

 

*The nozzle position was adjusted as far to the right as possible to provide a clear view of the vortex within the tube.

LC-MS/MS conditions

Agilent 1290 Infinity II LC system

  • 1290 Infinity II Multicolumn Thermostate, G7116B
  • 1290 Infinity II Multisampler, G7167B
  • 1290 Infinity II high speed pump, G7120A

Columns

  • InfinityLab PFC delay column, 4.6 x 30 mm, p/n 5062-8100
  • ZORBAX Eclipse Plus C18 guard column, 2.1 x 5 mm, 1.8 µm, p/n 821725-901
  • ZORBAX RRHD Eclipse Plus C18, 95 Å, 2.1 x 50 mm, 1.8 µm, p/n 959757-902

Mobile phases

A: 20 mM ammonium acetate in water

B: Methanol

 

Table 4. LC gradient

Time (min)

%A

%B

0.00

90

10

0.50

90

10

2.00

60

40

7.50

0

100

9.00

0

100

9.10

90

10

10.00

90

10

 

Flow rate: 0.4 mL/min

Injection volume: 5 µL

Column temperature: 50oC

Agilent 6470 MS/MS, G6470B

Gas temperature: 230oC

Gas flow: 4 L/min

Nebulizer: 20 psi

Sheath gas temperature: 375oC

Sheath gas flow: 12 L/min

Capillary voltage (positive): 3500 V

Capillary voltage (negative): 3500 V

Nozzle voltage (positive): 500 V

Nozzle voltage (negative): 0 V

Results

System calibration

For this application note, a total of five points were used in the calibration covering a range of 0.2 - 20 ng/L. The calibration curves were forced through zero and achieved excellent R2 values of greater than 0.999 for all target compounds. 

PFOS

Figure 1. Calibration curves for PFOS and PFOA. Calibration curves for the other target analytes in table 1 are shown in appendix C.

Determination of the minimum reporting level (MRL) and detection limit (DL)

A target MRL of 2 ng/L was selected, and eight replicate laboratory-fortified blanks (LFBs) were created at that concentration and extracted. The resulting data were used to calculate the DL for all target compounds. Figure 2 illustrates the calculated detection limit (DL) for all 30 ISO 21675 target compounds, ranging from 0.12 ng/L to 0.56 ng/L. Out of the 30 analyzed compounds, 26 had calculated DLs below 0.40 ng/L. One data point was excluded from each of the N-MeFOSA and N-EtFOSA datasets after being identified as an outlier. The statistical analysis was adjusted accordingly to account for one fewer data point in each dataset.

 

Figure 2. Target MRL and calculated DLs for the ISO 21675 target analytes. Compounds marked with an asterisk were analyzed in salt form.

The data for individual compounds are shown in appendix D.

Initial demonstration of precision and accuracy (IDP and IDA)

Four LFB samples were prepared at a concentration of 10 ng/L. The resulting data were used to determine the precision and accuracy of the sample preparation process. Figure 3 shows accuracy, while figure 4 shows precision. Excluding N-MeFOSA and N-EtFOSA, all target compounds exhibited mean recoveries between 91% and 121% of the spiked amount. Calculated CVs were 11.7% or lower for all compounds. N-MeFOSA and N-EtFOSA had mean recoveries of 57% and 55%, respectively, and are known to exhibit lower recoveries due to being neutral precursors that differ from the predominantly ionic compounds included in this method. Lower and more variable recoveries for these compounds have been documented in ISO 21675, suggesting that this behavior may be analyte-specific rather than extraction-specific. Overall, these results compare favorably with those reported in ISO 21675.

Figure 3. Initial demonstration of accuracy (10 ng/L, n=4). Compounds marked with an asterisk were analyzed in salt form.

 

Figure 4. Initial demonstration of precision (10 ng/L, n=4). Compounds marked with an asterisk were analyzed in salt form.

The data for individual compounds are shown in Appendix E.

Demonstration of low system background

A study was conducted to investigate each part of the extraction process and its potential contribution to PFAS background. This sequential process shown below identifies each component of the full extraction process that was evaluated for its potential PFAS background contribution. The components were then combined in the complete workflow used to extract the reagent water blank samples.

Table 5. Sequential evaluation of potential PFAS background contributions

Test

Biotage® PrepXpert-8

EVOLUTE® PFAS

Collection tube

TurboVap® LV

LC-TQ

Solvent

LC-TQ blank

 

 

 

 

X

X

TurboVap® LV blank

 

 

X

X

X

X

Column blank

 

X

X

X

X

X

Biotage® PrepXpert-8 blank

X

 

X

X

X

X

Reagent water blank

X

X

X

X

X

X

The results of these tests are given in appendix F, and selected data are shown in figures 5-8.

Figure 5. Contribution of the TurboVap® LV to PFAS background. Compounds marked with an asterisk were analyzed in salt form.

Figure 6. Contribution of the EVOLUTE® PFAS SPE cartridges to PFAS background. Compounds marked with an asterisk were analyzed in salt form.

Figure 7. Contribution of the Biotage® PrepXpert-8 to PFAS background. Compounds marked with an asterisk were analyzed in salt form

Figure 8. PFAS background for laboratory reagent blanks (LRB) extraction using EVOLUTE® PFAS SPE cartridges. Compounds marked with an asterisk were analyzed in salt form.

For results generated using only the analytical system, all target analytes were reported as not detected (N.D.) and therefore not shown.

When examining the resulting data for the Biotage® PrepXpert-8, EVOLUTE® PFAS SPE cartridges, and TurboVap® LV blanks, only trace levels of PFAS were observed. These trace levels were below the lowest calibration level and were not considered significant enough to interfere with water sample extractions. When evaluating the LRB sample extractions, low PFAS background concentrations were observed. All measured concentrations were below one-third of the corresponding analyte-specific MRL, meeting the background acceptance criterion. One clear outlier was observed in the LRB results for PFBA in sample 1 at 0.18 ng/L. This was identified as a single incident, and no supporting data indicate elevated PFBA concentrations in the other samples. The 0.18 ng/L result remained below the lowest curve point of the calibration, indicating the cleanliness of the Biotage® PrepXpert-8.

Examination of system carryover

To simulate contaminated drinking water samples, four LFB samples were prepared at concentrations more than twice the highest level on the calibration curve. These samples were extracted, and the cleaning procedure given in appendix A was performed three times. To verify that the system background was adequately reduced, a set of four LRB samples was extracted immediately after the cleaning procedure and analyzed. The LRB data obtained from this study are presented in appendix F.

The trace levels of PFAS targets that were observed were non-detect for all compounds. These results show that the cleaning procedure given in appendix A reduced carryover in the Biotage® PrepXpert-8 system. However, if concentrations of any PFAS compounds remain above the acceptance criteria, the cleaning procedure should be repeated, or the use of alternative cleaning solvents should be examined to help re-establish the system background to acceptable levels. 

Conclusion

The results of this application note demonstrate that the Biotage® PrepXpert-8 system, in combination with EVOLUTE® PFAS SPE cartridges and the TurboVap® LV, provides a reliable and efficient platform for the extraction of the PFAS analytes outlined in ISO 21675. For 28 of the 30 target compounds, mean recoveries ranged from 91% to 121%. Calculated CVs of 11.7% or lower for all compounds. Calculated DLs ranged from 0.12 to 0.56 ng/L. The testing demonstrated low PFAS background, with all measured concentrations below the corresponding acceptance criteria. Overall, the evaluated workflow provides laboratories with a reliable, automated solution for PFAS sample preparation.

Literature number: AN1037