Increased awareness of PFAS contamination has driven the need for reliable analytical methods to support environmental monitoring initiatives. Due to their extensive historical use and environmental persistence, PFAS compounds are now routinely targeted in water quality assessments throughout the world. ISO 21675 provides a standardized protocol for the extraction and determination of selected PFAS compounds in water, promoting accuracy and comparability of results between laboratories. In this work, an automated workflow employing the Biotage® PrepXpert-8 system and EVOLUTE® PFAS cartridges was evaluated for compliance with ISO 21675 sample preparation and performance requirements. Following extraction, samples were concentrated using the TurboVap® LV platform, resulting in a streamlined and reproducible process well suited for high-throughput PFAS testing laboratories.
Table 1. Listing of ISO 21675 target analytes
|
Analyte name |
Acronym |
CAS |
|
ISO 21675 target compounds |
||
|
Perfluoro-n-butanoic acid |
PFBA |
375–22–4 |
|
Perfluoro-n-pentanoic acid |
PFPeA |
2706–90–3 |
|
Perfluoro-n-butanesulfonic acid |
PFBS |
375–73–5 |
|
Perfluoro-n-hexanoic acid |
PFHxA |
307–24–4 |
|
Hexafluoropropylene oxide dimer acid |
HFPO-DA (GenX) |
13252–13–6 |
|
Perfluoro-n heptanoic acid |
PFHpA |
375–85–9 |
|
Perfluoro-n hexanesulfonic acid |
PFHxS |
355–46–4 |
|
4,8-Dioxa-3-H-perfluorononanoic acid |
ADONA (DONA) |
919005–14–4 |
|
Perfluoro-n heptanesulfonic acid |
PFHpS |
375–92–8 |
|
6:2 Fluorotelomer sulfonic acid |
6:2 FTS |
27619–97–2 |
|
Perfluoro-n-octanoic acid |
PFOA |
335–67–1 |
|
Perfluoro-n-octanesulfonic acid |
PFOS |
1763–23–1 |
|
Perfluoro-n-nonanoic acid |
PFNA |
375–95–1 |
|
8:2 Fluorotelomer unsaturated carboxylic acid |
8:2 FTUCA |
70887-84-2 |
|
Perfluoro-n decanoic acid |
PFDA |
335–76–2 |
|
9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid |
9Cl-PF3ONS |
73606–19–6 |
|
8:2 Fluorotelomer sulfonic acid |
8:2 FTS |
39108–34–4 |
|
N-methyl perfluorooctanesulfonamidoacetic acid |
N-MeFOSAA |
2355-31-9 |
|
Perfluorooctanesulfonamide |
FOSA |
754–91–6 |
|
Perfluoro-n-decanesulfonic acid |
PFDS |
335–77–3 |
|
Perfluoro-n-undecanoic acid |
PFUnA (PFUnDA) |
2058–94–8 |
|
N-ethyl perfluorooctanesulfonamidoacetic acid |
N-EtFOSAA |
2991-50-6 |
|
Perfluoro-n-dodecanoic acid |
PFDoA (PFDoDA) |
307–55–1 |
|
Perfluoro-n-tridecanoic acid |
PFTrDA |
72629–94–8 |
|
N-methyl perfluorooctanesulfonamide |
N-MeFOSA |
31506-32-8 |
|
N-ethyl perfluorooctanesulfonamide |
N-EtFOSA |
4151-50-2 |
|
Perfluoro-n-tetradecanoic acid |
PFTeDA |
376-06-7 |
|
Perfluoro-n-hexadecanoic acid |
PFHxDA |
67905-19-5 |
|
8:2 Polyfluoroalkyl phosphate diester |
8:2 diPAP |
678-41-1 |
|
Perfluoro-n-octadecanoic acid |
PFOcDA |
16517-11-6 |
Table 2. Extraction parameters
|
Operation |
Amount (mL or time) |
Input |
Output |
Flow (mL/min) |
|
Condition |
5 |
0.1% NH4OH in MeOH |
Solvent waste |
5 |
|
Condition |
5 |
MeOH |
Solvent waste |
5 |
|
Condition |
5 |
Water |
Water waste |
5 |
|
Load |
265 |
Sample |
Water waste |
10 |
|
Wash |
5 |
Water |
Water waste |
5 |
|
Purge |
5 |
Air |
Water waste |
5 |
|
Dry |
5 min |
Nitrogen |
Water waste |
- |
|
Rinse |
5 |
0.1% NH4OH in MeOH |
Sample |
60 |
|
Wait |
15 sec |
|||
|
Elute |
8 |
Sample |
Vial 2 |
5 |
|
Rinse |
5 |
0.1% NH4OH in MeOH |
Sample |
60 |
|
Wait |
15 sec |
|||
|
Elute |
8 |
Sample |
Vial 2 |
5 |
|
Purge |
5 |
Air |
Vial 2 |
5 |
Table 3. TurboVap® LV concentration protocol
|
Bath temperature: |
60 °C |
|
Evaporation mode: |
Method (ramp gradient) |
|
Manifold setup: |
48 positions |
|
Rack row height: |
120 mm |
|
Step 1: |
1.5 L/min for 20 min |
|
Step 2: |
3.0 L/min for 15 min |
|
Step 3: |
3.5 L/min for 50 min |
*The nozzle position was adjusted as far to the right as possible to provide a clear view of the vortex within the tube.
A: 20 mM ammonium acetate in water
B: Methanol
Table 4. LC gradient
|
Time (min) |
%A |
%B |
|
0.00 |
90 |
10 |
|
0.50 |
90 |
10 |
|
2.00 |
60 |
40 |
|
7.50 |
0 |
100 |
|
9.00 |
0 |
100 |
|
9.10 |
90 |
10 |
|
10.00 |
90 |
10 |
Flow rate: 0.4 mL/min
Injection volume: 5 µL
Column temperature: 50oC
Gas temperature: 230oC
Gas flow: 4 L/min
Nebulizer: 20 psi
Sheath gas temperature: 375oC
Sheath gas flow: 12 L/min
Capillary voltage (positive): 3500 V
Capillary voltage (negative): 3500 V
Nozzle voltage (positive): 500 V
Nozzle voltage (negative): 0 V
For this application note, a total of five points were used in the calibration covering a range of 0.2 - 20 ng/L. The calibration curves were forced through zero and achieved excellent R2 values of greater than 0.999 for all target compounds.
A target MRL of 2 ng/L was selected, and eight replicate laboratory-fortified blanks (LFBs) were created at that concentration and extracted. The resulting data were used to calculate the DL for all target compounds. Figure 2 illustrates the calculated detection limit (DL) for all 30 ISO 21675 target compounds, ranging from 0.12 ng/L to 0.56 ng/L. Out of the 30 analyzed compounds, 26 had calculated DLs below 0.40 ng/L. One data point was excluded from each of the N-MeFOSA and N-EtFOSA datasets after being identified as an outlier. The statistical analysis was adjusted accordingly to account for one fewer data point in each dataset.
Figure 2. Target MRL and calculated DLs for the ISO 21675 target analytes. Compounds marked with an asterisk were analyzed in salt form.
The data for individual compounds are shown in appendix D.
Four LFB samples were prepared at a concentration of 10 ng/L. The resulting data were used to determine the precision and accuracy of the sample preparation process. Figure 3 shows accuracy, while figure 4 shows precision. Excluding N-MeFOSA and N-EtFOSA, all target compounds exhibited mean recoveries between 91% and 121% of the spiked amount. Calculated CVs were 11.7% or lower for all compounds. N-MeFOSA and N-EtFOSA had mean recoveries of 57% and 55%, respectively, and are known to exhibit lower recoveries due to being neutral precursors that differ from the predominantly ionic compounds included in this method. Lower and more variable recoveries for these compounds have been documented in ISO 21675, suggesting that this behavior may be analyte-specific rather than extraction-specific. Overall, these results compare favorably with those reported in ISO 21675.
The data for individual compounds are shown in Appendix E.
A study was conducted to investigate each part of the extraction process and its potential contribution to PFAS background. This sequential process shown below identifies each component of the full extraction process that was evaluated for its potential PFAS background contribution. The components were then combined in the complete workflow used to extract the reagent water blank samples.
Table 5. Sequential evaluation of potential PFAS background contributions
|
Test |
Biotage® PrepXpert-8 |
EVOLUTE® PFAS |
Collection tube |
TurboVap® LV |
LC-TQ |
Solvent |
|
LC-TQ blank |
|
|
|
|
X |
X |
|
TurboVap® LV blank |
|
|
X |
X |
X |
X |
|
Column blank |
|
X |
X |
X |
X |
X |
|
Biotage® PrepXpert-8 blank |
X |
|
X |
X |
X |
X |
|
Reagent water blank |
X |
X |
X |
X |
X |
X |
The results of these tests are given in appendix F, and selected data are shown in figures 5-8.
For results generated using only the analytical system, all target analytes were reported as not detected (N.D.) and therefore not shown.
When examining the resulting data for the Biotage® PrepXpert-8, EVOLUTE® PFAS SPE cartridges, and TurboVap® LV blanks, only trace levels of PFAS were observed. These trace levels were below the lowest calibration level and were not considered significant enough to interfere with water sample extractions. When evaluating the LRB sample extractions, low PFAS background concentrations were observed. All measured concentrations were below one-third of the corresponding analyte-specific MRL, meeting the background acceptance criterion. One clear outlier was observed in the LRB results for PFBA in sample 1 at 0.18 ng/L. This was identified as a single incident, and no supporting data indicate elevated PFBA concentrations in the other samples. The 0.18 ng/L result remained below the lowest curve point of the calibration, indicating the cleanliness of the Biotage® PrepXpert-8.
To simulate contaminated drinking water samples, four LFB samples were prepared at concentrations more than twice the highest level on the calibration curve. These samples were extracted, and the cleaning procedure given in appendix A was performed three times. To verify that the system background was adequately reduced, a set of four LRB samples was extracted immediately after the cleaning procedure and analyzed. The LRB data obtained from this study are presented in appendix F.
The trace levels of PFAS targets that were observed were non-detect for all compounds. These results show that the cleaning procedure given in appendix A reduced carryover in the Biotage® PrepXpert-8 system. However, if concentrations of any PFAS compounds remain above the acceptance criteria, the cleaning procedure should be repeated, or the use of alternative cleaning solvents should be examined to help re-establish the system background to acceptable levels.
The results of this application note demonstrate that the Biotage® PrepXpert-8 system, in combination with EVOLUTE® PFAS SPE cartridges and the TurboVap® LV, provides a reliable and efficient platform for the extraction of the PFAS analytes outlined in ISO 21675. For 28 of the 30 target compounds, mean recoveries ranged from 91% to 121%. Calculated CVs of 11.7% or lower for all compounds. Calculated DLs ranged from 0.12 to 0.56 ng/L. The testing demonstrated low PFAS background, with all measured concentrations below the corresponding acceptance criteria. Overall, the evaluated workflow provides laboratories with a reliable, automated solution for PFAS sample preparation.
Literature number: AN1037