Automated PFAS sample preparation for ISO 21675
Introduction
Increased awareness of PFAS contamination has driven the need for reliable analytical methods to support environmental monitoring initiatives. Due to their extensive historical use and environmental persistence, PFAS compounds are now routinely targeted in water quality assessments throughout the world. ISO 21675 provides a standardized protocol for the extraction and determination of selected PFAS compounds in water, promoting accuracy and comparability of results between laboratories. In this work, an automated workflow employing the Biotage® PrepXpert-8 system and EVOLUTE® PFAS cartridges was evaluated for compliance with ISO 21675 sample preparation and performance requirements. Following extraction, samples were concentrated using the TurboVap® LV platform, resulting in a streamlined and reproducible process well suited for high-throughput PFAS testing laboratories.
Analytes
Table 1. Listing of ISO 21675 target analytes
|
Analyte name |
Acronym |
CAS |
|
ISO 21675 target compounds |
||
|
Perfluoro-n-butanoic acid |
PFBA |
375–22–4 |
|
Perfluoro-n-pentanoic acid |
PFPeA |
2706–90–3 |
|
Perfluoro-n-butanesulfonic acid |
PFBS |
375–73–5 |
|
Perfluoro-n-hexanoic acid |
PFHxA |
307–24–4 |
|
Hexafluoropropylene oxide dimer acid |
HFPO-DA (GenX) |
13252–13–6 |
|
Perfluoro-n heptanoic acid |
PFHpA |
375–85–9 |
|
Perfluoro-n hexanesulfonic acid |
PFHxS |
355–46–4 |
|
4,8-Dioxa-3-H-perfluorononanoic acid |
ADONA (DONA) |
919005–14–4 |
|
Perfluoro-n heptanesulfonic acid |
PFHpS |
375–92–8 |
|
6:2 Fluorotelomer sulfonic acid |
6:2 FTS |
27619–97–2 |
|
Perfluoro-n-octanoic acid |
PFOA |
335–67–1 |
|
Perfluoro-n-octanesulfonic acid |
PFOS |
1763–23–1 |
|
Perfluoro-n-nonanoic acid |
PFNA |
375–95–1 |
|
8:2 Fluorotelomer unsaturated carboxylic acid |
8:2 FTUCA |
70887-84-2 |
|
Perfluoro-n decanoic acid |
PFDA |
335–76–2 |
|
9-Chlorohexadecafluoro-3-oxanonane-1-sulfonic acid |
9Cl-PF3ONS |
73606–19–6 |
|
8:2 Fluorotelomer sulfonic acid |
8:2 FTS |
39108–34–4 |
|
N-methyl perfluorooctanesulfonamidoacetic acid |
N-MeFOSAA |
2355-31-9 |
|
Perfluorooctanesulfonamide |
FOSA |
754–91–6 |
|
Perfluoro-n-decanesulfonic acid |
PFDS |
335–77–3 |
|
Perfluoro-n-undecanoic acid |
PFUnA (PFUnDA) |
2058–94–8 |
|
N-ethyl perfluorooctanesulfonamidoacetic acid |
N-EtFOSAA |
2991-50-6 |
|
Perfluoro-n-dodecanoic acid |
PFDoA (PFDoDA) |
307–55–1 |
|
Perfluoro-n-tridecanoic acid |
PFTrDA |
72629–94–8 |
|
N-methyl perfluorooctanesulfonamide |
N-MeFOSA |
31506-32-8 |
|
N-ethyl perfluorooctanesulfonamide |
N-EtFOSA |
4151-50-2 |
|
Perfluoro-n-tetradecanoic acid |
PFTeDA |
376-06-7 |
|
Perfluoro-n-hexadecanoic acid |
PFHxDA |
67905-19-5 |
|
8:2 Polyfluoroalkyl phosphate diester |
8:2 diPAP |
678-41-1 |
|
Perfluoro-n-octadecanoic acid |
PFOcDA |
16517-11-6 |
Solution preparation
Ammonia methanol solution
- Add 400 μL of NH4OH to every 99.6 mL of methanol in a clean beaker.
- Mix thoroughly to homogenize the solution.
- Prepare fresh solution daily before extraction.
Sample preparation procedure
- Clean the Biotage® PrepXpert-8 automated SPE system using the procedure given in appendix A. If the cleanliness of a particular channel is in question, it is recommended to repeat the cleaning procedure and verify channel cleanliness by extracting blanks.
- Set up and fill new sample containers with water.
- Adjust the sample pH to 3 using glacial acetic acid.
- Spike the extracted labeled standards into each sample at a concentration corresponding to the midpoint of the calibration curve.
- If necessary to perform statistical studies, spike a known amount of PFAS target standard into the water samples to achieve the desired concentration.
- Attach each prepared water sample container to the sample bottle cap on the Biotage® PrepXpert-8 system and ensure that the sip tube is positioned to drain all liquid from the sample bottle.
- Attach the EVOLUTE® PFAS SPE cartridges to the adapters of the Biotage® PrepXpert-8 system.
- Load 15 mL centrifuge tubes onto the collection vessel rack and ensure that the rack is installed on the Biotage® PrepXpert-8 automated tray.
- Program in the parameters for the extraction of samples by ISO 21675 given in table 2, adjust the sample-load volume as required, and run the method. The protocol will result in an approximately 10 mL extract.
- Determine the initial sample volume by either measuring the volume required to refill the sample container to the original mark using a graduated cylinder or by weighing the container before and after sample loading.
- Transfer the centrifuge tubes to the TurboVap® LV system and concentrate the extract to a volume of less than 0.5 mL using nitrogen according to the parameters in Table 3.
- Add methanol to each extract to achieve a final volume of 0.5 – 1.0 mL.
- Add an appropriate amount of internal standard to each extract and mix to homogenize.
- Inject the extract into a calibrated LC-MS/MS system and analyze it using the conditions given in the following sections.
Table 2. Extraction parameters
|
Operation |
Amount (mL or time) |
Input |
Output |
Flow (mL/min) |
|
Condition |
5 |
0.1% NH4OH in MeOH |
Solvent waste |
5 |
|
Condition |
5 |
MeOH |
Solvent waste |
5 |
|
Condition |
5 |
Water |
Water waste |
5 |
|
Load |
265 |
Sample |
Water waste |
10 |
|
Wash |
5 |
Water |
Water waste |
5 |
|
Purge |
5 |
Air |
Water waste |
5 |
|
Dry |
5 min |
Nitrogen |
Water waste |
- |
|
Rinse |
5 |
0.1% NH4OH in MeOH |
Sample |
60 |
|
Wait |
15 sec |
|||
|
Elute |
8 |
Sample |
Vial 2 |
5 |
|
Rinse |
5 |
0.1% NH4OH in MeOH |
Sample |
60 |
|
Wait |
15 sec |
|||
|
Elute |
8 |
Sample |
Vial 2 |
5 |
|
Purge |
5 |
Air |
Vial 2 |
5 |
Table 3. TurboVap® LV concentration protocol
|
60 °C |
|
|
Evaporation mode: |
Method (ramp gradient) |
|
Manifold setup: |
48 positions |
|
Rack row height: |
120 mm |
|
Step 1: |
1.5 L/min for 20 min |
|
Step 2: |
3.0 L/min for 15 min |
|
Step 3: |
3.5 L/min for 50 min |
*The nozzle position was adjusted as far to the right as possible to provide a clear view of the vortex within the tube.
LC-MS/MS conditions
Agilent 1290 Infinity II LC system
- 1290 Infinity II Multicolumn Thermostate, G7116B
- 1290 Infinity II Multisampler, G7167B
- 1290 Infinity II high speed pump, G7120A
Columns
- InfinityLab PFC delay column, 4.6 x 30 mm, p/n 5062-8100
- ZORBAX Eclipse Plus C18 guard column, 2.1 x 5 mm, 1.8 µm, p/n 821725-901
- ZORBAX RRHD Eclipse Plus C18, 95 Å, 2.1 x 50 mm, 1.8 µm, p/n 959757-902
Mobile phases
A: 20 mM ammonium acetate in water
B: Methanol
Table 4. LC gradient
|
Time (min) |
%A |
%B |
|
0.00 |
90 |
10 |
|
0.50 |
90 |
10 |
|
2.00 |
60 |
40 |
|
7.50 |
0 |
100 |
|
9.00 |
0 |
100 |
|
9.10 |
90 |
10 |
|
10.00 |
90 |
10 |
Flow rate: 0.4 mL/min
Injection volume: 5 µL
Column temperature: 50oC
Agilent 6470 MS/MS, G6470B
Gas temperature: 230oC
Gas flow: 4 L/min
Nebulizer: 20 psi
Sheath gas temperature: 375oC
Sheath gas flow: 12 L/min
Capillary voltage (positive): 3500 V
Capillary voltage (negative): 3500 V
Nozzle voltage (positive): 500 V
Nozzle voltage (negative): 0 V
Results
System calibration
For this application note, a total of five points were used in the calibration covering a range of 0.2 - 20 ng/L. The calibration curves were forced through zero and achieved excellent R2 values of greater than 0.999 for all target compounds.
PFOS
Figure 1. Calibration curves for PFOS and PFOA. Calibration curves for the other target analytes in table 1 are shown in appendix C.
Determination of the minimum reporting level (MRL) and detection limit (DL)
A target MRL of 2 ng/L was selected, and eight replicate laboratory-fortified blanks (LFBs) were created at that concentration and extracted. The resulting data were used to calculate the DL for all target compounds. Figure 2 illustrates the calculated detection limit (DL) for all 30 ISO 21675 target compounds, ranging from 0.12 ng/L to 0.56 ng/L. Out of the 30 analyzed compounds, 26 had calculated DLs below 0.40 ng/L. One data point was excluded from each of the N-MeFOSA and N-EtFOSA datasets after being identified as an outlier. The statistical analysis was adjusted accordingly to account for one fewer data point in each dataset.

Figure 2. Target MRL and calculated DLs for the ISO 21675 target analytes. Compounds marked with an asterisk were analyzed in salt form.
The data for individual compounds are shown in appendix D.
Initial demonstration of precision and accuracy (IDP and IDA)
Four LFB samples were prepared at a concentration of 10 ng/L. The resulting data were used to determine the precision and accuracy of the sample preparation process. Figure 3 shows accuracy, while figure 4 shows precision. Excluding N-MeFOSA and N-EtFOSA, all target compounds exhibited mean recoveries between 91% and 121% of the spiked amount. Calculated CVs were 11.7% or lower for all compounds. N-MeFOSA and N-EtFOSA had mean recoveries of 57% and 55%, respectively, and are known to exhibit lower recoveries due to being neutral precursors that differ from the predominantly ionic compounds included in this method. Lower and more variable recoveries for these compounds have been documented in ISO 21675, suggesting that this behavior may be analyte-specific rather than extraction-specific. Overall, these results compare favorably with those reported in ISO 21675.
Figure 3. Initial demonstration of accuracy (10 ng/L, n=4). Compounds marked with an asterisk were analyzed in salt form.
Figure 4. Initial demonstration of precision (10 ng/L, n=4). Compounds marked with an asterisk were analyzed in salt form.
The data for individual compounds are shown in Appendix E.
Demonstration of low system background
A study was conducted to investigate each part of the extraction process and its potential contribution to PFAS background. This sequential process shown below identifies each component of the full extraction process that was evaluated for its potential PFAS background contribution. The components were then combined in the complete workflow used to extract the reagent water blank samples.
Table 5. Sequential evaluation of potential PFAS background contributions
|
Test |
Biotage® PrepXpert-8 |
EVOLUTE® PFAS |
Collection tube |
TurboVap® LV |
LC-TQ |
Solvent |
|
LC-TQ blank |
|
|
|
|
X |
X |
|
TurboVap® LV blank |
|
|
X |
X |
X |
X |
|
Column blank |
|
X |
X |
X |
X |
X |
|
Biotage® PrepXpert-8 blank |
X |
|
X |
X |
X |
X |
|
Reagent water blank |
X |
X |
X |
X |
X |
X |
The results of these tests are given in appendix F, and selected data are shown in figures 5-8.
Figure 5. Contribution of the TurboVap® LV to PFAS background. Compounds marked with an asterisk were analyzed in salt form.
Figure 6. Contribution of the EVOLUTE® PFAS SPE cartridges to PFAS background. Compounds marked with an asterisk were analyzed in salt form.
Figure 7. Contribution of the Biotage® PrepXpert-8 to PFAS background. Compounds marked with an asterisk were analyzed in salt form
Figure 8. PFAS background for laboratory reagent blanks (LRB) extraction using EVOLUTE® PFAS SPE cartridges. Compounds marked with an asterisk were analyzed in salt form.
For results generated using only the analytical system, all target analytes were reported as not detected (N.D.) and therefore not shown.
When examining the resulting data for the Biotage® PrepXpert-8, EVOLUTE® PFAS SPE cartridges, and TurboVap® LV blanks, only trace levels of PFAS were observed. These trace levels were below the lowest calibration level and were not considered significant enough to interfere with water sample extractions. When evaluating the LRB sample extractions, low PFAS background concentrations were observed. All measured concentrations were below one-third of the corresponding analyte-specific MRL, meeting the background acceptance criterion. One clear outlier was observed in the LRB results for PFBA in sample 1 at 0.18 ng/L. This was identified as a single incident, and no supporting data indicate elevated PFBA concentrations in the other samples. The 0.18 ng/L result remained below the lowest curve point of the calibration, indicating the cleanliness of the Biotage® PrepXpert-8.
Examination of system carryover
To simulate contaminated drinking water samples, four LFB samples were prepared at concentrations more than twice the highest level on the calibration curve. These samples were extracted, and the cleaning procedure given in appendix A was performed three times. To verify that the system background was adequately reduced, a set of four LRB samples was extracted immediately after the cleaning procedure and analyzed. The LRB data obtained from this study are presented in appendix F.
The trace levels of PFAS targets that were observed were non-detect for all compounds. These results show that the cleaning procedure given in appendix A reduced carryover in the Biotage® PrepXpert-8 system. However, if concentrations of any PFAS compounds remain above the acceptance criteria, the cleaning procedure should be repeated, or the use of alternative cleaning solvents should be examined to help re-establish the system background to acceptable levels.
Conclusion
The results of this application note demonstrate that the Biotage® PrepXpert-8 system, in combination with EVOLUTE® PFAS SPE cartridges and the TurboVap® LV, provides a reliable and efficient platform for the extraction of the PFAS analytes outlined in ISO 21675. For 28 of the 30 target compounds, mean recoveries ranged from 91% to 121%. Calculated CVs of 11.7% or lower for all compounds. Calculated DLs ranged from 0.12 to 0.56 ng/L. The testing demonstrated low PFAS background, with all measured concentrations below the corresponding acceptance criteria. Overall, the evaluated workflow provides laboratories with a reliable, automated solution for PFAS sample preparation.
Literature number: AN1037