Extraction of cocaine and metabolites from whole blood using ISOLUTE® SLE+
By Biotage
Figure 1. Structure of Benzoylecgonine
Introduction
This application note describes the extraction of cocaine and major metabolites from whole blood, prior to GC/MS analysis. This protocol also allows the simultaneous extraction of various other drugs of abuse classes: amphetamines, barbiturates, benzodiazepines and opiates ISOLUTE® SLE+ cartridges with 1 mL sample capacity are used to extract whole blood samples following a straightforward sample dilution. No protein precipitation or other pre-treatment is required prior to sample loading. The sample preparation procedure delivers clean extracts, good recoveries and RSD values and LLOQs from 20 ng/mL (analyte dependant).
ISOLUTE® SLE+ Supported Liquid Extraction plates and cartridges offer an efficient alternative to traditional liquid- liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation, and significantly reduced sample preparation.
Analytes
Anhydroecgonine methyl ester (AEME), Ecgonine (EME), Cocaine, Cocaethylene, Benzoylecgonine-D3, (BZE-D3), Benzoylegonine (BZE)
Sample preparation procedure
Format
ISOLUTE® SLE+ 1 mL Sample Volume cartridge, part number 820-0140-C
Sample pre-treatment
To 1 mL of whole blood, add 10 µL of ISTD (total 100 ng/mL). Allow to equilibrate and add 1 mL of 1% ammonium hydroxide (aq). Vortex.
Sample loading
Load 750 μL of the pre-treated whole blood onto the cartridge and apply a pulse of vacuum or positive pressure (3–5 seconds) to initiate flow. Allow the sample to absorb for 5 minutes.
Analyte extraction
Apply dichloromethane (DCM, 2.5 mL) and allow to flow under gravity for 5 minutes. Collect in an appropriate glass tube.
Apply a second aliquot of DCM (2.5 mL) and allow to flow under gravity for 5 minutes. Apply vacuum or positive pressure (5–10 seconds) to pull through any remaining extraction solvent into the collection vessel.
Post-elution and reconstitution
Evaporate the extract in a stream of air or nitrogen using a TurboVap LV (ambient, 20 to 40 L/min).
Reconstitute the extracts with ethyl acetate (250 µL) and vortex for 20 seconds before transferring to high recovery GC vials. Evaporate the extract in a stream of air or nitrogen using a SPE Dry (40 °C, 20 to 40 L/min).
Reconstitute extracts with ethyl acetate (40 µL) and BSTFA (with 1% t-BDMCS) (40 µL), vortex and heat for 30 minutes at 70 °C to complete derivatization.
GC Conditions
Instrument
Agilent 7890A with QuickSwap
Column
Agilent J&W DB-5, 30 m x 0.25 mm ID x 0.25 μm
Carrier
Helium 1.2 mL/min (constant flow)
Inlet
260 °C, Splitless, purge flow: 50 mL/min at 1.0 min
Injection
2 μL
Wash solvents
Acetone & ethyl acetate
Oven
Initial temperature 60 °C, hold for 1 minute, ramp 50 °C/min to 200 °C , hold for 1.5 minutes, ramp 10 °C/min to 250°C.
Post-run
Backflush for 1.6 minutes (2 void volumes)
Transfer line
280 °C
MS conditions
Instrument
Agilent 5975C
Source
230 °C
Quadrupole
150 °C
MSD mode
SIM
SIM parameters
|
SIM Group |
Analyte |
Target (Quant) Ion |
1st Qual Ion |
2nd Qual Ion |
|
1 |
AEME |
152 |
181 |
|
|
2 |
EME |
82 |
96 |
|
|
3 |
Cocaine |
94 |
82 |
|
|
4 |
Cocaethylene |
196 |
82 |
94 |
|
4 |
BZE-D3 |
85 |
243 |
|
|
4 |
BZE |
82 |
240 |
|
Results
Blank whole blood was spiked at 100 ng/mL for recovery testing. Reproducible data was observed from the typical recovery data with RSD values <10% as shown in Figure 2.
Figure 2. Typical recoveries for cocaine and metabolites.
Figure 3. Total Ion Chromatogram of cocaine and metabolites at 100 ng/mL
Calibration curves
Whole blood was spiked prior to extraction, at concentrations of 10, 20, 50, 75, 100, 200 and 500 ng/mL for each analyte to create calibration curves. BZE-D3 was spiked at 100 ng/mL for each level. The curves are shown in Figure 4.
Figure 4. Charts demonstrating coefficient of determination
(r2) values between 0.9953 and 0.9995 for the cocaine and metabolites
|
Drug Analyte |
DCM LLOQ (ng/mL) |
|---|---|
|
AEME |
20 |
|
EME |
50 |
|
Cocaine |
50 |
|
Cocaethylene |
50 |
|
BZE |
50 |
Additional notes
Solvents and reagent preparation:
- All solvents were HPLC grade.
- 1% ammonium hydroxide (aq): Add concentrated ammonium hydroxide (28–30%) (1 mL) to HPLC grade water (99 mL).
Cartridge loading: ISOLUTE® SLE+ cartridges are underloaded (750 µL sample on a 1 mL capacity cartridge) to avoid breakthrough of whole blood matrix.
Simultaneous extraction of other drug classes: This protocol allows the simultaneous extraction of amphetamines, barbiturates, benzodiazepines and opiates.
Ordering information
|
Part Number |
Description |
Quantity |
|
820-0140-C |
ISOLUTE® SLE+ 1 mL Sample Volume Cartridge* |
30 |
|
820-0140-CG |
ISOLUTE SLE+ 1 mL Sample Volume Cartridge (tabless) |
30 |
|
PPM-48 |
Biotage® PRESSURE+ 48 Positive Pressure Manifold |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry Sample Concentrator System 220/240 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry Sample Concentrator System 100/120 V |
1 |
Literature number: AN856