Extraction of opiates from whole blood using ISOLUTE® SLE+

By Biotage

 

biotage-an852-figure1-structure-morphine
Figure 1. Structure of Morphine


Introduction


This application note describes the extraction of opiate compounds from whole blood, prior to GC/MS analysis. This protocol also allows the simultaneous extraction of various other drugs of abuse classes: amphetamines, barbiturates, benzodiazepines and cocaine ISOLUTE® SLE+ cartridges with 1 mL sample capacity are used to extract whole blood samples following a straightforward sample dilution. No protein precipitation or other pre-treatment is required prior to sample loading. The sample preparation procedure delivers clean extracts, good recoveries and RSD values and LLOQs from 50 ng/mL (analyte dependant).


ISOLUTE® SLE+ Supported Liquid Extraction plates and cartridges offer an efficient alternative to traditional liquid- liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation, and significantly reduced sample preparation.

Analytes


Dihydrocodeine, hydrocodone, codeine, hydromorphone, oxycodone, morphine-D3, morphine, 6-monoacetylmorphine (6-MAM)-D3, 6-MAM, oxymorphone.
 

Sample preparation procedure


Format:


ISOLUTE® SLE+ 1mL Sample Volume cartridge, part number 820-0140-C


Sample pre-treatment


To 1 mL of whole blood, add 10 µL of ISTD (total 100 ng/mL). Allow to equilibrate and add 1 mL of 1% ammonium hydroxide (aq). Vortex.


Sample loading


Load 750 μL of the pre-treated whole blood onto the cartridge and apply a pulse of vacuum or positive pressure (3–5 seconds) to initiate flow. Allow the sample to absorb for 5 minutes.


Analyte extraction


Apply dichloromethane (DCM, 2.5 mL) and allow to flow under gravity for 5 minutes. Collect in an appropriate glass tube. Apply a second aliquot of DCM* (2.5 mL) and allow to flow under gravity for 5 minutes. Apply vacuum or positive pressure (5–10 seconds) to pull through any remaining extraction solvent into the collection vessel.


*Note that MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, benzodiazepines, barbiturates). If simultaneous extraction of cocaine and its BZE metabolite is required, DCM should be used as extraction solvent.


Post-elution and reconstitution


Evaporate the extract in a stream of air or nitrogen using a TurboVap® LV (ambient, 20 to 40 L/min). Reconstitute the extracts with ethyl acetate (250 µL) and vortex for 20 seconds before transferring to high recovery GC vials.


Evaporate the extract in a stream of air or nitrogen using a Biotage® SPE Dry (40 °C, 20 to 40 L/min). Reconstitute extracts with ethyl acetate (40 µL) and BSTFA (with 1% t-BDMCS) (40 µL), vortex and heat on a block for 30 minutes at 70 °C to complete derivatization.

 

GC conditions


Instrument


Agilent 7890A with QuickSwap


Column


Agilent J&W DB-5, 30 m x 0.25 mm ID x 0.25 μm


Carrier


Helium 1.2 mL/min (constant flow)


Inlet


250 °C, Splitless, purge flow: 50 mL/min at 1.0 min


Injection


1 µL


Wash solvents


Acetone & ethyl acetate


Oven


Initial temperature 80 °C, hold for 1 minute, ramp 10 °C/min to 273 °C


Post run


Backflush for 1.6 minutes (2 void volumes)


Transfer line


280°C
 

MS conditions


Instrument


Agilent 5975C


Source


230 °C


Quadrupole


150 °C


MSD mode


SIM


SIM parameters

SIM Group

Analyte

Target (Quant) Ion

Qual Ion 1

Qual Ion 2

1

Dihydrocodeine

373

146

 

2

Hydrocodone

178

371

196

2

Codeine

115

299

242

3

Hydromorphone

357

300

 

3

Oxycodone

372

273

212

3

Morphine-D3

432

239

 

3

Morphine

429

236

 

4

6-MAM-D3

402

343

 

4

6-MAM

399

340

 

5

Oxymorphone

445

288

 

 

Results


Blank whole blood was spiked at 100 ng/mL for recovery testing; the typical recovery data is shown in Figure 2. Both MTBE and DCM protocols gave reproducible data with RSD values <10%.

biotage-an852-figure2-typical-opiate-recoveriesFigure 2. Typical opiate recoveries using MTBE or DCM as extraction solvent.  

biotage-an852-figure3-total-ion-chromatogram-optiatesFigure 3. Total Ion Chromatogram of optiates at 100 ng/mL using the DCM
extraction protocol.

 

Calibration curves


Whole blood was spiked prior to extraction, at concentrations of 10, 20, 50, 75, 100, 200 and 500 ng/mL for each analyte to create calibration curves. Morphine-D3 and 6-MAM-D3 were spiked at 100 ng/mL for each level. The curves are shown in Figure 4.

biotage-an852-figure4-linearity-demonstrating-coefficient-determinationFigure 4. Linearity demonstrating coefficient of determination (r2) values between 0.9967 and 0.9998 for the opiates using the DCM protocol.

 

Additional notes


Solvents and reagent preparation:

  • All solvents were HPLC grade.
  • 1% ammonium hydroxide (aq): Add concentrated ammonium hydroxide (28-30%) (1 mL) to HPLC grade water (99 mL).


Cartridge loading:

ISOLUTE® SLE+ cartridges are underloaded (750 µL sample on a 1 mL capacity cartridge) to avoid breakthrough of whole blood matrix.


Non-chlorinated extraction solvent alternative:

MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, benzodiazepines, barbiturates). If simultaneous extraction of cocaine and its BZE metabolite is required, DCM should be used as extraction solvent.

Ordering information

Part Number

Description

Quantity

820-0140-C

ISOLUTE® SLE+ 1 mL Sample Volume Cartridge*

30

820-0140-CG

ISOLUTE SLE+ 1 mL Sample Volume Cartridge (tabless)

30

PPM-48

Biotage® PRESSURE+ 48 Positive Pressure Manifold

1

SD-9600-DHS-EU

Biotage® SPE Dry Sample Concentrator System 220/240 V

1

SD-9600-DHS-NA

Biotage® SPE Dry Sample Concentrator System 100/120 V

1

 

Literature number: AN852