Fast extraction of metanephrine and normetanephrine from pooled plasma using EVOLUTE® EXPRESS WCX
By Biotage
This application note describes a method of extraction for metanephrine and normetanephrine from pooled plasma using EVOLUTE® EXPRESS WCX 96-well SPE plates.

Figure 1. Structures metanephrine and normetanephrine
Introduction
The normalized metabolite concentrations of epinephrine and norepinephrine precursors have common utility in clinical diagnostics related to neuroendocrine tumours. Of particular interest are the selected metabolites metanephrine and normetanephrine. The challenges in population screening for these metabolites are time, cost and efficacy of sample preparation strategies required for trace level determinations (pg/mL-ng/mL levels) by liquid chromatography-tandem mass spectrometry. A new solid phase extraction technology has been developed to reduce the time associated with sample preparation methods by eliminating traditional steps in the extraction workflow. Leveraging this technology, a method protocol was developed using a load, wash and elute approach. This method was optimized for pooled plasma.
EVOLUTE EXPRESS WCX ion exchange SPE materials offer an efficient alternative to traditional ion exchange SPE plates for bioanalytical sample preparation, providing high analyte recoveries and significantly reduced sample preparation time.
Analytes
Metanephrine, normetanephrine
Sample preparation procedure
Format: EVOLUTE EXPRESS WCX 30 mg 96-well plate, part number 602-0030-PXO1
Matrix: Pooled plasma
Sample Pre-treatment: Dilute plasma sample with ammonium acetate (3:1 v/v, 0.05 M, pH = 7)
Sample Loading: Load 0.4 mL of the diluted sample into each well. Apply positive pressure (PRESSURE+ 96 Positive Pressure Manifold, PPM-96) to maintain a steady flow rate of 1 mL/min (10-12 drops)
Wash 1*: Remove interferences with water (1 mL)
Wash 2*: Remove interferences with methanol/acetonitrile (1:1 v/v, 1 mL)
Elution: Elute analytes with methanol/acetonitrile (1:1 v/v):formic acid (95:5, v/v)
Blow down/recon: Evaporate eluate to dryness and reconstitute in 0.4 mL of mobile phase
*Suggested variable to try on implementation: evaluate 2x wash steps for additional extract cleanliness.
HPLC conditions
Instrument: Agilent 1200 Liquid Handling System (Agilent Technologies, Berkshire, UK)
Cartridge: Restek Organic Acids 150 mm x 4.6 mm (5 µm) (catalog # 9165565)
Mobile Phase: Solvent A: 1% Formic acid in water
Solvent B: Acetonitrile
Flow rate: 1 mL /min
Injection Volume: 100 μL
Temperature: 35 °C
|
Step |
Time (min) |
Flow Rate (mL/min) |
%A |
%B |
|
1 2 |
0.0 3.0 |
1.0 1.0 |
90 90 |
10 10 |
Mass spectrometry conditions
Instrument: Applied Biosystems/MDS Sciex 4000 Q-Trap triple quadrupole mass
spectrometer (Applied Biosystems, Foster City, CA.) equipped with a Turbo Ionspray® interface for mass analysis.
Ion Source Temperature: 600 °C
|
Scan Function |
Analyte |
MRM Transition |
Declustering Potential (DP) |
Collision Energy (CE) |
Cell Exit Potential (CXP) |
|
1 2 |
Metanephrine Normetanephrine |
180 → 148 166 → 134 |
25 25 |
20 20 |
16 16 |

Figure 2. A typical ion chromatogram for plasma fortified at 150 ng/mL

Figure 3. Relative recovery of metanephrine and normetanephrine from fortified plasma using an EVOLUTE® EXPRESS WCX 30 mg 96-well plate. Repeatability as %RSD was <15% for both analytes.
Ordering information
|
Part Number |
Description |
Quantity |
|
602-0030-PX01 |
EVOLUTE® EXPRESS WCX 30 mg Fixed Well Plate |
1 |
|
PPM-96 |
Biotage® PRESSURE+ 96 Positive pressure Manifold 96 well |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry Sample Concentrator System 220 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry Sample Concentrator System 110 V |
1 |
Literature number: AN794