Extraction of benzodiazepines from whole blood using ISOLUTE® SLE+

By Biotage

biotage-an854-figure1-structure-diazepamFigure 1. Structure of Diazepam

Introduction


This application note describes the extraction of benzodiazepine compounds from whole blood, prior to GC/MS analysis. This protocol also allows the simultaneous extraction of various other drugs of abuse classes: amphetamines, barbiturates, cocaine and opiates.
ISOLUTE® SLE+ cartridges with 1 mL sample capacity are used to extract whole blood samples following a straightforward sample dilution. No protein precipitation or other pre-treatment is required prior to sample loading. The sample preparation procedure delivers clean extracts, good recoveries and RSD values and LLOQs from 10 ng/mL (analyte dependant).


ISOLUTE® SLE+ Supported Liquid Extraction plates and cartridges offer an efficient alternative to traditional liquid- liquid extraction (LLE) for bioanalytical sample preparation, providing high analyte recoveries, no emulsion formation, and significantly reduced sample preparation.

Analytes


Diazepam-D5, Diazepam, Nordiazepam, Midazolam, Flunitrazepam, 7-amino-flunitrazepam, Bromazepam, Oxazepam, Nitrazepam, Flurazepam, Temazepam,
7-amino-clonazepam, Lorazepam, Clonazepam, 2-hydroxy- ethyl-flurazepam, Estazolam, Alprazolam, Triazolam, Alpha- hydroxy-alprazolam, Alpha-hydroxy-triazolam
 

Sample preparation procedure


Format:


ISOLUTE® SLE+ 1mL Sample Volume cartridge, part number 820-0140-C


Sample pre-treatment


To 1 mL of whole blood, add 10 µL of ISTD (total 100 ng/mL). Allow to equilibrate and add 1 mL of 1% ammonium hydroxide (aq). Vortex.


Sample loading


Load 750 μL of the pre-treated whole blood onto the cartridge and apply a pulse of vacuum or positive pressure (3–5 seconds) to initiate flow. Allow the sample to absorb for 5 minutes


Analyte extraction


Apply dichloromethane* (DCM, 2.5 mL) and allow to flow under gravity for 5 minutes. Collect in an appropriate glass tube. Apply a second aliquot of DCM* (2.5 mL) and allow to flow under gravity for 5 minutes. Apply vacuum or positive pressure (5–10 seconds) to pull through any remaining extraction solvent into the collection vessel.


* Note that MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, barbiturates, opiates). If simultaneous extraction of cocaine and metabolites is required, DCM should be used as extraction solvent.


Post-elution and reconstitution


Evaporate the extract in a stream of air or nitrogen using a TurboVap® LV (40 °C, 20 to 40 L/min). Reconstitute the extracts with ethyl acetate (250 µL) and vortex for 20 seconds before transferring to high recovery GC vials.
Evaporate the extract in a stream of air or nitrogen using a SPE Dry (40 °C, 20 to 40 L/min). Reconstitute extracts with ethyl acetate (50 µL) and MTBSTFA (with 1% t-BDMCS) (50 µL), vortex and heat on a block for 30 minutes at 70 °C to complete derivatization.

 

GC conditions


Instrument


Agilent 7890A with QuickSwap


Column


Agilent J&W DB-5, 30 m x 0.25 mm ID x 0.25 μm


Carrier


Helium 1.2 mL/min (constant flow)


Inlet


260 °C, Splitless, purge flow: 50 mL/min at 1.0 min


Injection


1 μL


Wash solvents


Acetone & ethyl acetate


Oven


Initial temperature 60 °C, hold for 1 minute, ramp 12 °C/min to
330 °C, hold for 0.5 minutes


Post-run


Backflush for 1.6 minutes (2 void volumes)


Transfer Line


280 °C

MS Conditions


Instrument


Agilent 5975C


Source


230 °C


Quadrupole


150 °C


MSD mode


SIM
 

SIM parameters

SIM Group

Analyte

Target (Quant) Ion

Qual Ion

1

Diazepam-D5

261

289

1

Diazepam

256

284

2

Nordiazepam

327

329

3

Midazolam

310

312

3

Flunitrazepam

312

285

3

7-amino-flunitrazepam

255

283

3

Bromazepam

372

374

4

Oxazepam

457

291

4

Nitrazepam

338

291

4

Flurazepam

86

99

5

Temazepam

357

359

6

7-amino-clonazepam

342

344

6

Lorazepam

491

513

6

Clonazepam

372

374

6

2-hydroxy-ethyl-flurazepam

389

390

6

Estazolam

294

205

7

Alprazolam

279

273

8

Triazolam

313

238

9

Alpha-hydroxy-alprazolam

381

383

10

Alpha-hydroxy-triazolam

415

417

Results


Blank whole blood was spiked at 100 ng/mL for recovery testing; the typical recovery data is shown in Figure 2. Both MTBE and DCM protocols gave reproducible data with RSD values <10%.
biotage-an854-figure2-typical-benzodiazepine-recoveries-using-MTBE-or-DCM
Figure 2. Typical benzodiazepine recoveries using MTBE or DCM as extraction solvent.

biotage-an854-figure3-Total-ion-chromatogram-application-analytesFigure 3. Total Ion Chromatogram of application analytes at 100 ng/mL using the DCM extraction protocol
 

Calibration curves


Whole blood was spiked prior to extraction at concentrations of 10, 20, 50, 75, 100, 200 and 500 ng/mL for each analyte to create calibration curves. Diazepam-D5 was spiked at 100 ng/mL for each level. The curves are shown in Figure 4.

biotage-an854-figure4-chart-demonstrating-r2biotage-an854-figure5-chart-demonstrating-r2

Figure 4. Chart demonstrating coefficient of determination (r2) values between 0.9922 and
0.9999 for benzodiazepines using the DCM extraction protocol.

 

Drug Analyte  

LLOQ (ng/mL) DCM Extraction

LLOQ (ng/mL) MTBE Extraction

Diazepam

20

10

Nordiazepam

<10

<10

Midazolam

20

20

Flunitrazepam

75

50

7-amino-flunitrazepam

100

100

Bromazepam

50

50

Oxazepam

10

20

Nitrazepam

50

50

Flurazepam

20

20

Temazepam

10

20

7-amino-clonazepam

50

100

Lorazepam

20

20

Clonazepam

75

75

2-hydroxy-ethyl-flurazepam

20

50

Estazolam

100

100

Alprazolam

75

75

Triazolam

100

100

Alpha-hydroxy-alprazolam

20

20

Alpha-hydroxy-triazolam

50

50

 

Additional notes

Solvents and reagent preparation:

  • All solvents were HPLC grade.
  • 1% ammonium hydroxide (aq): Add concentrated ammonium hydroxide (28–30%) (1 mL) to HPLC grade water (99 mL).


Cartridge loading: ISOLUTE® SLE+ cartridges are underloaded (750 µL sample on a 1 mL capacity cartridge) to avoid breakthrough of whole blood matrix.
Non-chlorinated extraction solvent alternative: MTBE can be used as an alternative extraction solvent if a non-chlorinated option is required. MTBE also suitable for extraction of other analyte classes (amphetamines, barbiturates, opiates). If simulta- neous extraction of cocaine and metabolites is required, DCM should be used as extraction solvent.


Ordering information

Part Number

Description

Quantity

820-0140-C

ISOLUTE® SLE+ 1 mL Sample Volume Cartridge*

30

820-0140-CG

ISOLUTE SLE+ 1 mL Sample Volume Cartridge (tabless)

30

PPM-48

Biotage® PRESSURE+ 48 Positive Pressure Manifold

1

SD-9600-DHS-EU

Biotage® SPE Dry Sample Concentrator System 220/240 V

1

SD-9600-DHS-NA

Biotage® SPE Dry Sample Concentrator System 100/120 V

1

 

Literature number: AN854

Published: Dec 7, 2025 11:45:00 AM