Dec 8, 2025 11:17:22 AM
Extraction of a comprehensive steroid panel from human serum using EVOLUTE® EXPRESS ABN
By Biotage
For research use only. NOT for use in diagnostic procedures.
Figure 1. Structures of DHEAS, Estradiol and Testosterone.
Introduction
This application note describes the extraction of a panel of 19 steroid hormones from human serum using EVOLUTE® EXPRESS ABN solid phase extraction plates prior to LC/MS analysis. The simple sample preparation procedure delivers clean extracts and analyte recoveries greater than 80% with RSDs lower than 10% for the majority of analytes. Linearity of >0.99 is achieved for all analytes in the range 5–5000 pg/mL. Manual sample preparation was performed using the Biotage® Pressure+ 96 Positive Pressure Manifold. The sample preparation method is automatable using the Biotage® ExtraheraTM. See Appendix for automation parameters and comparative data generated using the automated method.
EVOLUTE EXPRESS products dramatically improve flow characteristics, and enhance sample preparation productivity.
Analytes
Cortisol, 18-OH-Corticosterone, 21-Deoxycortisol, Cortisone, Estradiol, 17-OH-Pregnenolone, Aldosterone, 11-Deoxycortisol, Corticosterone, Estrone, Dihydroepiandrosterone (DHEA), 17-OH-Progesterone, Dehydroepiandrosterone Sulfate (DHEAS), Testosterone, Dihydrotestosterone (DHT), Pregnenolone, Androstenedione, 11-Deoxycorticosterone, Progesterone
Internal standards
DHT-D3 and Aldosterone-D4
Sample preparation procedure
Format
EVOLUTE® EXPRESS ABN 10 mg plate, part number 600-0010-PX01.
Sample pre-treatment
Spike serum with methanolic internal standard solution (25 µL/1 mL serum).
Dilute pre-spiked serum (200 µL) with 1% formic acid (aq) (200 µL) and mix.
Condition (optional)
Condition wells with methanol (500 µL).
Equilibration (optional)
Equilibrate wells with 0.1% formic acid (aq) (500 µL).
Sample loading
Load 400 µL of pre-treated serum into each well.
Wash 1
Elute interferences with H2O (500 µL).
Wash 2
Elute interferences with H2O:MeOH (60:40, v/v, 500 µL).
Elution
For panels without DHEAS, elute analytes with ethyl acetate (150 µL).
For panels including DHEAS, elute analytes with methanol (150 µL).
Post elution and reconstitution
Dry the extract in a stream of air or nitrogen using a Biotage® SPE Dry at 40 °C, 20 to 40 L/min for 20 minutes.
Reconstitute evaporated samples with 50:50 mobile phase: A mobile phase B (200 µL) and mix thoroughly.
UHPLC conditions
Instrument
Shimadzu Nexera X2 UHPLC
Column
ACE C18 (100 mm x 2.1 mm, 1.7 µm)
(Advanced Chromatography Technologies Ltd, Aberdeen, UK) with EXP Guard column holder fitted with a C-18 cartridge (Restek, UK)
Mobile phase
A: 0.2 mM ammonium fluoride (aq)
B: Methanol
Flow rate
0.4 mL min
Column temperature
40 oC
Injection volume
10 µL
Table 1. UHPLC gradient.
|
Time (min) |
%A |
%B |
|
0 |
50 |
50 |
|
3 |
40 |
60 |
|
8 |
10 |
90 |
|
9 |
5 |
95 |
|
9.1 |
5 |
95 |
|
9.5 |
50 |
50 |
Mass spectrometry conditions
Instrument
Shimadzu 8060 Triple Quadrupole MS using ES interface
Nebulizing gas flow
3 L/min
Drying gas flow
3 L/min
Heating gas flow
17 L/min
Interface temperature
400 oC
DL temperature
250 oC
Heat block temperature
400 oC
CID gas flow
270 kPa
Table 2. MS conditions for target analytes in positive and negative mode.
|
Analytes |
MRM Transition |
Collision Energy |
Ion Mode |
|
DHEAS |
367.1 > 97.05 |
33 |
- |
|
Cortisol |
363.4 > 121.25 |
-24 |
+ |
|
18-OH-Corticosterone |
363.3 > 269.2 |
-16 |
+ |
|
Cortisone |
361.3 > 163.15 |
-22 |
+ |
|
21-Deoxycortisol |
347.1 > 311.2 |
-16 |
+ |
|
Estradiol |
271.1 > 145.2 |
39 |
- |
|
Aldosterone-D4 |
363.1 > 190.3 |
19 |
- |
|
Aldosterone |
359.1 > 189.25 |
18 |
- |
|
17-OH-Pregnenolone |
315.3 > 297.2 |
-13 |
+ |
|
11-Deoxycortisol |
347.3 > 109.25 |
-27 |
+ |
|
Corticosterone |
347.3 > 329.25 |
-16 |
+ |
|
Estrone |
269.2 > 145.2 |
37 |
- |
|
11-Deoxycorticosterone |
331.3 > 109.05 |
-25 |
+ |
|
DHEA |
289.3 > 253.2 |
-13 |
+ |
|
Testosterone |
289.3 > 97.05 |
-23 |
+ |
|
DHT-D3 |
294.4 > 258.25 |
-16 |
+ |
|
DHT |
291.3 > 255.25 |
-15 |
+ |
|
Androstenedione |
287.3 > 97.2 |
-21 |
+ |
|
Pregnenolone |
299.3 > 159.25 |
-20 |
+ |
|
17-OH-Progesterone |
331.3 > 97.1 |
-22 |
+ |
|
Progesterone |
315.2 > 97.2 |
-22 |
+ |
Results
If the polar analyte DHEAS is to be included in the steroid suite, methanol should be used as the elution solvent. If DHEAS is not required, an alternative elution solvent of ethyl acetate can be used. Recovery data for both elution solvent systems is shown below in Figure 2. Note: Due to the significant levels of these endogenous analytes, recovery data was generated using stripped serum as matrix. The optimized SPE protocols delivered analyte recoveries above 80% for the majority of analytes, with corresponding RSDs below 10%.
Removal of phospholipids was demonstrated by monitoring MRM transitions using the common phospholipid product ion at m/z 184. Figure 3 shows the serum phospholipid profile comparing lysophospholipids (bottom trace) and larger molecular weight phospholipids (top trace) extracted using the optimized SPE protocols and 100 µL of serum protein precipitated with 400 µL of acetonitrile. The use of a water immiscible organic solvent such as ethyl acetate eliminated phospholipids compared to traditional elution solvents such as methanol or acetonitrile.
Recoveries
Figure 2. Typical analyte % extraction recoveries (n=7) using MeOH or EtOAc as elution solvents.
Figure 3. Phospholipid profile comparison between protein precipitated serum and optimized SPE extracts using MeOH or EtOAc as elution solvents.
Figure 4 shows representative chromatography obtained from stripped serum spiked at 5 ng/mL. Satisfactory resolution of the various isobars was obtained using the ACE C18 UHPLC column. In order to achieve low level detection of analytes in positive and negative ion modes a combination of 0.2 mM NH4F (aq) and MeOH was utilized as mobile phase.
Linearity was investigated using stripped serum spiked between 5–5000 pg/mL. Good linearity was observed for all analytes, typically delivering r2 values greater than 0.99. Table 3 details linearity performance and associated LOQ for each analyte using MeOH and EtOAc as elution solvents. Selected calibration curves using EtOAc as elution solvent are demonstrated in Figure 5.
Figure 4. Representative chromatography for stripped serum spiked at 5 ng/mL.
|
Analytes |
r2 MeOH |
LLOQ (pg/mL) MeOH |
r2 EtOAc |
LLOQ (pg/mL) EtOAc |
|
DHEAS |
0.991 |
25 |
- |
- |
|
Cortisol |
0.992 |
< 50 |
0.997 |
< 50 |
|
18-OH-Corticosterone |
0.996 |
< 250 |
0.990 |
< 250 |
|
Cortisone |
0.996 |
< 5 |
0.994 |
< 5 |
|
21-Deoxycortisol |
0.995 |
50 |
0.996 |
50 |
|
Estradiol |
0.997 |
50 |
0.996 |
25 |
|
Aldosterone |
0.976 |
< 25 |
0.990 |
< 25 |
|
17-OH-Pregnenolone |
0.996 |
500 |
0.990 |
500 |
|
11-Deoxycortisol |
0.993 |
25 |
0.997 |
10 |
|
Corticosterone |
0.992 |
100 |
0.996 |
100 |
|
Estrone |
0.996 |
< 25 |
0.997 |
10 |
|
11-Deoxycorticosterone |
0.999 |
< 10 |
0.997 |
< 10 |
|
DHEA |
0.994 |
< 1000 |
0.995 |
< 1000 |
|
Testosterone |
0.997 |
< 5 |
0.997 |
< 5 |
|
DHT |
0.998 |
< 250 |
0.990 |
< 250 |
|
Androstenedione |
0.997 |
25 |
0.998 |
10 |
|
Pregnenolone |
0.994 |
< 250 |
0.985 |
< 250 |
|
17-OH-Progesterone |
0.995 |
< 25 |
0.993 |
< 25 |
|
Progesterone |
0.995 |
10 |
0.998 |
10 |

Figure 5. Calibration curves for Estradiol (a), Testosterone (b), 17-OH-Progesterone (c) and Androstenedione (d).
Chemicals and reagents
- Standards (including deuterated internal standards) and ammonium fluoride (LCMS grade) were purchased from Sigma-Aldrich Company Ltd. (Gillingham, UK).
- Formic acid, 98% methanol and ethyl acetate (all LCMS grade) were purchased from Honeywell Research Chemicals (Bucharest, Romania).
- Water (18.2 MΩ.cm) was drawn fresh daily from a Direct-Q 5 water purifier (Merck Millipore, Watford, UK).
- Pooled human plasma was obtained from The Welsh Blood Service (Pontyclun, UK).
- Stripped serum was purchased from Golden West Biologicals, Inc. (Temecula CA).
- 0.1% formic acid was prepared by adding 100 µL of formic acid (as above) to 99.9 mL of water.
- H2O:MeOH (60:40, v/v) was prepared by adding 60 mL water to 40 mL methanol, and mixing.
- 0.2 mM ammonium fluoride was prepared by adding 7.4 mg of ammonium fluoride (as above) to 1 L of water, and mixing.
- Internal standard preparation: 10 µL of a 10 pg/µL stock solution (DHT-D3 and Aldosterone-D4) was added to 990 µL of methanol.
Additional information
- Ammonium fluoride increased sensitivity in both positive and negative ion modes.
- Other strategies for increasing sensitivity:
- Increase matrix volumes above 200 µL .
- Decrease reconstitution solvent volume below 200 µL.
- Increase injection volumes above 10 µL.
- The addition of IPA into EtOAc up to 10% can increase recovery of aldosterone using this method.
- Steroids can exhibit non-specific binding to plastic collection plates. Different plastics exhibit different binding characteristics. Addition of 2 µL of ethylene glycol to the collection plate prior to evaporation can mitigate this issue. Note: No ethylene glycol was used in generation of the data shown in this application note, utilizing collection plate p/n 121-5203.
- To minimize the number of solvent lines required when automating this method using Biotage ® Extrahera™, the equilibration solvent can be replaced with 1% formic acid (as per sample pre-treatment step).
- 1% formic acid was prepared by adding 1.0 mL of formic acid(as above) to 99.0 mL of water.
Ordering information
|
Part Number |
Description |
Quantity |
|
600-0010-PX01 |
EVOLUTE® EXPRESS ABN 10 mg Plate |
1 |
|
SD-9600-DHS-EU |
Biotage® SPE Dry 96 Sample Evaporator 220/240 V |
1 |
|
SD-9600-DHS-NA |
Biotage® SPE Dry 96 Sample Evaporator 100/120 V |
1 |
|
121-5203 |
Collection Plate, 2 mL Square |
50 |
|
PPM-96 |
Biotage® PRESSURE+ 96 Positive Pressure Manifold |
1 |
Appendix: Biotage® Extrahera™ settings
The method described in this application note was automated on the Biotage® Extrahera™ using EVOLUTE® Express ABN 10 mg plates. This appendix contains the software settings required to configure Extrahera to run this method using ethyl acetate as elution solvent, for panels NOT including DHEAS is shown). Results generated using the automated method show enhanced analyte recovery and lower RSD compared to the manual processing method. In addition, reduced LLOQ and improved linearity were achieved. Comparative data examples are shown on page 11.
|
Sample Name: |
Steroid Express ABN – EtAOc elution |
|
Sample Plate/Rack: |
Plate |
|
Extraction Media: |
EVOLUTE® EXPRESS ABN 10 mg |






Solvent properties
|
Solvent Description |
|
|
1 |
Methanol |
|
2 |
0.1% Formic acid |
|
3 |
Deionized water |
|
4 |
MeOH:H2O 40:60 |
|
5 |
Ethyl Acetate |
|
6 |
|
|
7 |
|
|
8 |
|
|
9 |
|
|
10 |
|
Solvent |
1 |
2 |
3 |
4 |
5 |
6 |
7 |
8 |
9 |
10 |
|
Reservoir Type |
Refillable |
Non Refillable |
||||||||
|
Capacity |
||||||||||
|
Aspiration flow rate (mL/min) |
10 |
10 |
10 |
10 |
10 |
|||||
|
Dispense flow rate (mL/min) |
20 |
20 |
20 |
20 |
10 |
|||||
|
Lower air gap flow rate (mL/min) |
20 |
20 |
20 |
20 |
20 |
|||||
|
Lower air gap volume (µL) |
5 |
5 |
5 |
5 |
5 |
|||||
|
Upper air gap flow rate (mL/min) |
120 |
20 |
20 |
20 |
120 |
|||||
|
Upper air gap volume (µL) |
100 |
100 |
100 |
100 |
100 |
|||||
|
Upper air gap dispense pause |
300 |
300 |
300 |
300 |
300 |
|||||
|
Conditioning? |
Yes |
Yes |
Yes |
Yes |
Yes |
|||||
|
Conditioning number of times |
3 |
3 |
3 |
3 |
3 |
|||||
|
Conditioning flow rate (mL/min) |
20 |
20 |
20 |
20 |
10 |
|||||
|
Chlorinated |
No |
No |
No |
No |
No |
|||||
|
Serial dispense |
No |
No |
No |
No |
No |
|||||




Results
Extrahera automated method:
Figure 6 (below) shows enhanced recovery and reduced % RSD achieved using the automated method compared to the manual approach. Table 4 shows improved linearity and reduced LLOQ are achieved with the automated method.
Figure 6. Steroid recoveries (Biotage® ExtraheraTM automated method vs manual method) with methanol elution (DHEAS included). % RSDs shown as error bars.
Table 4. Analyte linearity and LOQ performance (Ethyl acetate elution, DHEAS not included).
|
Analyte |
r2 EtOAc |
LLOQ (pg/mL) EtOAc |
Analyte |
r2 EtOAc |
LLOQ (pg/mL) EtOAc |
|
DHEAS |
|
|
Estrone |
0.997 |
10 |
|
Cortisol |
0.998 |
< 10 |
11-Deoxycorticosterone |
0.998 |
5 |
|
18-OH-Corticosterone |
0.997 |
50 |
DHEA |
0.999 |
250 |
|
Cortisone |
0.996 |
< 5 |
Testosterone |
0.997 |
<5 |
|
21-Deoxycortisol |
0.997 |
10 |
DHT |
0.994 |
100 |
|
Estradiol |
0.997 |
< 25 |
Androstenedione |
0.998 |
5 |
|
Aldosterone |
0.995 |
25 |
Pregnenolone |
0.997 |
< 250 |
|
17-OH-Pregnenolone |
0.994 |
< 100 |
17-OH-Progesterone |
0.995 |
5 |
|
11-Deoxycortisol |
0.997 |
5 |
Progesterone |
0.999 |
5 |
|
Corticosterone |
0.999 |
10 |
|
|
|
Literature Number: AN891
Published: Dec 8, 2025 11:17:22 AM