Biotage® Extrahera™ HV-5000

Automated sample preparation workstation

Automates high-volume SPE and SLE sample prep with SMART pipetting, GLP compliance, and intuitive software. 

Biotage-Extrahera-hv-5000-portrait-no-sw

Explore the main benefits of Biotage® Extrahera™ HV-5000

Biotage-Extrahera-hv-5000-portrait-sw
  • Pipette pump capacity

    250-5000 μL
  • Pipetting specifications

    For the 5000 μL pipette tip (P/N 417007):

    At 250 μL: ± 2.0% accuracy and 3.0% CV
    At 500 μL: ±1.5% accuracy and 2.0% CV
    At 1000 μL: ±1.0% accuracy and 1.5% CV
    At 5000 μL: ±0.5% accuracy and 1.0% CV

  • Pressure unit

    Positive pressure unit:
    Processing: 0-5 Bar (0.1 Bar Increments)
    Plate dry: 5 Bar at 600 mL/min

  • Format

    48-position extraction racks for 3 mL (tabless) columns

    24-position extraction racks for 3 and 6 mL (tabless) columns

    12-position extraction racks for 10 mL (tabless) columns

    12-position extraction racks for 15 mL (tabless) columns

  • Plate positions

    6 position upper deck used for sample, extraction plates, pipette tips and solvents.

    4 positions on lower deck carousel for multiple elution and waste.

  • Disposable tips

    Disposable sample tips (Position 2): 5000 μL clear tips

    Disposable solvent tips (Position 1): 5000 μL clear tips

    Disposable solvent tips (Position 6): 5000 μL clear tips

  • Temperature

    Operating temperature: 18°C to 32°C.

    Storage and transportation temperature: -25 °C to 60 °C

  • Humidity

    10 to 90% RH
  • Power supply

    100 to 240 V, 50/60 Hz
  • Fuses

    T4A at the power inlet (2 required)
  • Maximum power consumed

    300 VA
  • Pressurized air supply

    >5 L/min (0.18 cubic feet/min), 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi)
  • Inert gas supply

    AIR port requirements: >5 L/min (0.18 cubic feet/min), 6 ± 0.2 bar (0.6 ± 0.02 MPa; 87 ± 3 psi).

    N2 port requirements: 1 L/min (0.04 cubic feet/min) during processing and 70 L/min (2.5 cubic feet/min) during plate dry, 6 ± 0.2 bar (0.6 ± 0.02 MPa, 87 ± 3 psi)

  • Vacuum supply

    Partial pressure range: < 500 mbar (50 kPa; 7 psi)

    Flow rate: ≥ 8.3 L/min (0.29 cubic feet/min)

  • Max sound level

    65 dB(A)
  • GLP

    GLP software ready
  • Certifications

    CE, CSA certified

Resources

  • Application guide

    Guide to environmental applications using Biotage® Extrahera™ HV-5000

    Read article
  • Application note

    Automated extraction of 11-Nor-9-Carboxy-THC from human urine with ISOLUTE® HAX SPE columns using Biotage® Extrahera™ HV-5000, prior to GC-MS analysis

    Read article
  • Application note

    Automated extraction of Tetrahydrocannabinol (THC) and its metabolites from human whole blood with ISOLUTE® HAX SPE columns using Biotage® Extrahera™ HV-5000, prior to GC-MS analysis

    Read article
  • Application note

    Automated extraction of Amphetamines from human urine using ISOLUTE® HCX SPE columns with Biotage® Extrahera™ HV-5000, prior to GC-MS analysis

    Read article
  • Application note

    Automated extraction of a drugs of abuse panel from urine using ISOLUTE® HCX SPE columns with the Biotage® Extrahera™ HV-5000

    Read article
  • Application note

    Automated extraction of acrylamide from instant coffee using Biotage® Extrahera™ HV-5000, according to EN 16618:2015

    Read article
  • Application note

    Extraction of veterinary growth promoters from animal urine using ISOLUTE® SLE+ prior to UPLC-MS/MS analysis

    Read article
  • Application note

    Extraction of a comprehensive steroid panel from horse hair using ISOLUTE® SLE+ prior to LC/MS-MS analysis

    Read article
  • Application note

    Automated extraction of a drugs of abuse panel from human urine using ISOLUTE® SLE+ with Biotage® Extrahera™ HV-5000

    Read article

Ordering information

Part no.
Name
Pack size
Price
Part no.
417002
Name
Biotage® Extrahera™ HV-5000
Pack size
1
Price
Get a quote
Part no.
SER-EX-EXW
Name
Extended Warranty - 12 Month Extension - Biotage® Extrahera™
Pack size
1
Price
Get a quote
Part no.
SER-EX-FYMP
Name
First Year Maintenance Package - Biotage® Extrahera™
Pack size
1
Price
Get a quote
Part no.
SER-EX-FYMP-Q
Name
FYMP and Qualification Package (IQOQ) - Biotage® Extrahera™
Pack size
1
Price
Get a quote
Part no.
SER-EX-GLP-IQOQ
Name
GLP Qualification Package (IQOQ)
Pack size
1
Price
Get a quote
Part no.
SER-EX-IN
Name
Installation and Familiarization - Biotage® Extrahera™
Pack size
1
Price
Get a quote
Part no.
SER-EX-IQOQ
Name
Qualification Package (IQ/OQ) - Biotage® Extrahera™
Pack size
1
Price
Get a quote

Related products

Biotage® Lysera | Bead mill homogenizer

Designed for automating the grinding, lysing, and homogenizing of biological, food and environmental samples prior to sample preparation.

TurboVap® | Automated solvent evaporation systems

With four configurations of the same platform —TurboVap® LV, II, Eh and P+ are tailored to fit any sample preparation workflow.

ISOLUTE® cSPE for QuEChERS | Cartridges for no-shake QuEChERS clean-up

Prepacked cartridges designed for efficient clean-up of pesticide residues in fruit and vegetables.

ISOLUTE® HAX | Solid-phase extraction cartridges and plates

A mixed-mode sorbent providing clean extracts for acidic compounds from complex biological fluid samples using a dual retention mechanism.

Get a quote for Biotage® Extrahera™ HV-5000